EUVIMEDEUVIMED entdecken
ValidierungsstudieOpen Access · CC BY 4.0

Development and validation of a stability-indicating HPLC method for assay of tonabersat in pharmaceutical formulations

Santosh Bhujbal, Ilva D Rupenthal, Priyanka Agarwal

Methods · 2024

DOI: 10.1016/j.ymeth.2024.10.001 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift Methods, 2024. Verfasst von Santosh Bhujbal und 2 weiteren Autor:innen. Beteiligte Einrichtung: Buchanan Ocular Therapeutics Unit, Department of Ophthalmology, Aotearoa-New Zealand National Eye Centre, Faculty of Medical and Health Sciences, The…. Hauptthemen laut MEDLINE: Drug Stability. Die Arbeit wurde bisher 31-mal zitiert (Crossref) und nennt 50 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Health Research Council of New Zealand, Buchanan Charitable Foundation, Maurice and Phyllis Paykel Trust.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

A stability-indicating reversed-phase high-performance liquid chromatography (RP-HPLC) method was developed to assay tonabersat and assess its stability in pharmaceutical formulations. Chromatographic separation was achieved using a Kinetex® C18 column (2.6 µm, 150 x 3 mm, 100 Å) at 50 °C, with a 20 µL injection volume. A linear gradient of acetonitrile in water (5 - 33.5 %) was applied for 1 min, followed by a gradual increase to 100 % over 26 min at a flow rate of 0.5 mL/min. Tonabersat and its degradation products were detected at 275 nm and 210 nm, respectively. The optimized method was used to evaluate the stability of tonabersat in lipid-based pharmaceutical formulations at 5 ± 3 °C, 25 ± 2°C/60 ± 5 % RH, and 40 ± 2 °C/75 ± 5 % RH over 3 months. The method was validated as per ICH guidelines and demonstrated linearity in the range of 5 - 200 µg/mL (R2 = 0.99994) with good accuracy (98.25 - 101.58 % recovery) and precision (% RSD < 2.5 %). The limits of detection and quantitation were 0.8 µg/mL and 5 µg/mL, respectively. Forced degradation studies showed significant degradation on exposure to alkaline (90.33 ± 0.80 %), acidic (70.60 ± 1.57 %), and oxidative stress (33.95 ± 0.69 %) at 70 °C, but no degradation was observed on exposure to thermal or photolytic stress. No chemical degradation was observed in either formulation on storage. Thus, the method was sensitive, specific, and suitable for stability testing of tonabersat in pharmaceutical formulations.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Drug StabilityChromatography, High Pressure LiquidChromatography, Reverse-PhaseReproducibility of ResultsLimit of DetectionChemistry, Pharmaceutical

Stichwörter der Autor:innen

PharmaceuticalReversed-phase HPLCStability-indicatingTonabersatValidation

Autor:innen und Institutionen

  • Buchanan Ocular Therapeutics Unit, Department of Ophthalmology, Aotearoa-New Zealand National Eye Centre, Faculty of Medical and Health Sciences, The University of Auckland, Auckland 1142, New Zealand.
  • Buchanan Ocular Therapeutics Unit, Department of Ophthalmology, Aotearoa-New Zealand National Eye Centre, Faculty of Medical and Health Sciences, The University of Auckland, Auckland 1142, New Zealand. Electronic addres…

Förderung

  • Health Research Council of New Zealand
  • Buchanan Charitable Foundation
  • Maurice and Phyllis Paykel Trust

Zitieren

APA

Bhujbal, S., Rupenthal, I. D., & Agarwal, P. (2024). Development and validation of a stability-indicating HPLC method for assay of tonabersat in pharmaceutical formulations. Methods. https://doi.org/10.1016/j.ymeth.2024.10.001

Vancouver

Bhujbal S, Rupenthal ID, Agarwal P. Development and validation of a stability-indicating HPLC method for assay of tonabersat in pharmaceutical formulations. Methods. 2024. doi:10.1016/j.ymeth.2024.10.001

BibTeX
@article{Bhujbal2024,
  author = {Bhujbal, Santosh and Rupenthal, Ilva D and Agarwal, Priyanka},
  title = {Development and validation of a stability-indicating HPLC method for assay of tonabersat in pharmaceutical formulations},
  journal = {Methods},
  year = {2024},
  doi = {10.1016/j.ymeth.2024.10.001}
}
Kostenlos starten →