Performance of automated anterior segment OCT-based quantitative imaging in adult anterior chamber inflammation
Ophthalmology Science · 2026-11 · 101465
DOI: 10.1016/j.xops.2026.101465 ↗Kurz gefasst
Fachartikel in der Zeitschrift Ophthalmology Science, 2026-11. Verfasst von Ameenat L. Solebo und 25 weiteren Autor:innen. Die Arbeit nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Institute for Health and Care Research, NIHR Moorfields Biomedical Research Centre, Wellcome Trust.
Abstract (Originalsprache)
Objective
To investigate the performance of anterior segment (AS-) OCT quantitative imaging of anterior chamber inflammation in uveitis patients with diverse demographics.
Design
Prospective cross-sectional study.
Participants
144 adult patients managed at a tertiary care service in the UK.
Methods
Repeated swept-source AS-OCT imaging was performed pre- and post-pupil dilation (i.e. 4 scan sets). Inflammation was quantified using a validated human-in-the-loop automated image analysis pipeline, Minuscule Cell Detection (MCD), which identified and counted putative inflammatory cells on AS-OCT. Main Outcome Measures Test-retest variability of AS-OCT and diagnostic accuracy of various AS-OCT derived measurands (minimum, maximum, median counts per cross sectional image, and total counts across volume image sets per eye or MINCC, MAXCC, MEDCC and TOTCC) versus Standardization of Uveitis Nomenclature (SUN) grading system as assessed by a uveitis specialist.
Results
A total of 281 eyes were included in the analysis. Median age was 48 years (IQR 36-64). Strong test-retest measurand reliability was demonstrated, with a 95% tolerance interval ratio 0.3 - 3.0. The best diagnostic performances for SUN activity were observed with the MINCC threshold of 3 particles (negative predictive value for clinical activity of 89.8%, 95% CI 83.0 - 94.1). Associations between AS-OCT measurands and patient age (adjusted coefficient 7.5 additional particles, 95% CI 0.5 – 14.6, p
Förderung
- National Institute for Health and Care Research
- NIHR Moorfields Biomedical Research Centre
- Wellcome Trust
Zitieren
APA
Solebo, A. L., Chen, B., Aznan, N. K. N., Xochicale, M., Roberts, T., Petrushkin, H., Lim, C., Shu, R., Jacobson, M., Farisogullari, I., Demarinis, G., Abdelfattah, K., Tynan, D., Lotay, J., Vijan, K., Tsika, C., Williams, O., Clare, G., Testi, I., …, & Chu, C. J. (2026). Performance of automated anterior segment OCT-based quantitative imaging in adult anterior chamber inflammation. Ophthalmology Science, 101465. https://doi.org/10.1016/j.xops.2026.101465
Vancouver
Solebo AL, Chen B, Aznan NKN, Xochicale M, Roberts T, Petrushkin H, et al. Performance of automated anterior segment OCT-based quantitative imaging in adult anterior chamber inflammation. Ophthalmology Science. 2026:101465. doi:10.1016/j.xops.2026.101465
BibTeX
@article{Solebo2026,
author = {Solebo, Ameenat L. and Chen, Boyu and Aznan, Nik Khadijah Nik and Xochicale, Miguel and Roberts, Tom and Petrushkin, Harry and Lim, Claire and Shu, Rongling and Jacobson, Megan and Farisogullari, Idil and Demarinis, Giuseppe and Abdelfattah, Khaled and Tynan, Dominique and Lotay, Jivan and Vijan, Kanwaldeep and Tsika, Chrysanthi and Williams, Olayinka and Clare, Gerry and Testi, Ilaria and Tucker, William R. and Addison, Peter and Tsui, Edmund and Pavesio, Carlos and Rahi, Jugnoo and Taylor, Paul and Chu, Colin J.},
title = {Performance of automated anterior segment OCT-based quantitative imaging in adult anterior chamber inflammation},
journal = {Ophthalmology Science},
year = {2026},
pages = {101465},
doi = {10.1016/j.xops.2026.101465}
}
EUVIMED