Validation of the Reaction Time-Based False-Positive Rate as a Reliability Index for Standard Automated Perimetry

Hiroki Nomoto, Sachiko Okuyama, Chota Matsumoto, Tairo Kimura, Keiji Yoshikawa, Shiro Mizoue, Mami Nanno, Shinji Kimura, Satoshi Inoue, Shunji Kusaka

Ophthalmology Science · 2026-10 · 6 (10) · 101347

DOI: 10.1016/j.xops.2026.101347 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Ophthalmology Science, 2026-10. Verfasst von Hiroki Nomoto und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ophthalmology, Kindai University Faculty of Medicine, Osaka, Japan. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To validate a reaction time (RT)-based false-positive rate (RT-FP) for standard automated perimetry (SAP) and to determine the optimal RT listening window.

Design

Retrospective cross-sectional study.

Participants

A total of 1201 eyes from 607 glaucoma patients tested with the AIZE algorithm (imo perimeter, 24plus[1-2] pattern).

Methods

Reaction time-based FP rate was defined as the proportion of stimulus responses below a defined RT listening window, excluding catch trials. Listening windows of 100, 200, and 300 ms were compared using receiver operating characteristic (ROC) analysis against conventional catch-trial FP rate ≥15%. The effect of excluding responses at locations with estimated threshold ≤0, ≤5, ≤10, and ≤15 dB on the association between RT-FP and disease severity was evaluated using Spearman correlation with mean deviation (MD). Relative variability of each FP index was assessed using the coefficient of variation (CV). Main Outcome Measures: Area under the ROC curve (AUC) for discriminating conventional FP ≥15%, Spearman correlation with MD and CV.

Results

A 300-ms listening window achieved the highest discriminative accuracy (AUC = 0.783), outperforming 200-ms (AUC = 0.733) and 100-ms (AUC = 0.718) listening windows (both P 10 dB substantially reduced metric dependence on disease severity to levels comparable to conventional FP, with greater measurement stability and no requirement for dedicated catch trials. This optimized RT-FP represents a promising reliability index for SAP warranting prospective validation. Financial Disclosures: Proprietary or commercial disclosure may be found in the Footnotes and Disclosures at the end of this article.

Schlagwörter

Stichwörter der Autor:innen

AIZE algorithmFalse-positive rateReaction timeStandard automated perimetryVisual field

Autor:innen und Institutionen

  • Department of Ophthalmology, Kindai University Faculty of Medicine, Osaka, Japan.
  • Ueno Eye Clinic, Tokyo, Japan.
  • Yoshikawa Eye Clinic, Tokyo, Japan.
  • Department of Ophthalmology, Ehime University School of Medicine, Ehime, Japan.
  • Kagurasaka Minamino Eye Clinic, Tokyo, Japan.
  • CREWT Medical Systems, Inc., Tokyo, Japan.

Zitieren

APA

Nomoto, H., Okuyama, S., Matsumoto, C., Kimura, T., Yoshikawa, K., Mizoue, S., Nanno, M., Kimura, S., Inoue, S., & Kusaka, S. (2026). Validation of the Reaction Time-Based False-Positive Rate as a Reliability Index for Standard Automated Perimetry. Ophthalmology Science, 6(10), 101347. https://doi.org/10.1016/j.xops.2026.101347

Vancouver

Nomoto H, Okuyama S, Matsumoto C, Kimura T, Yoshikawa K, Mizoue S, et al. Validation of the Reaction Time-Based False-Positive Rate as a Reliability Index for Standard Automated Perimetry. Ophthalmology Science. 2026;6(10):101347. doi:10.1016/j.xops.2026.101347

BibTeX
@article{Nomoto2026,
  author = {Nomoto, Hiroki and Okuyama, Sachiko and Matsumoto, Chota and Kimura, Tairo and Yoshikawa, Keiji and Mizoue, Shiro and Nanno, Mami and Kimura, Shinji and Inoue, Satoshi and Kusaka, Shunji},
  title = {Validation of the Reaction Time-Based False-Positive Rate as a Reliability Index for Standard Automated Perimetry},
  journal = {Ophthalmology Science},
  year = {2026},
  volume = {6},
  number = {10},
  pages = {101347},
  doi = {10.1016/j.xops.2026.101347}
}
Kostenlos starten →