EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC-ND 4.0

Pivotal Trial Toward Effectiveness of Self-administered OCT in Neovascular Age-related Macular Degeneration. Report 2—Artificial Intelligence Analytics

Eric W Schneider, Jeffrey S Heier, Nancy M Holekamp, Miguel A Busquets, Alan L Wagner, S Krishna Mukkamala, Christopher D Riemann, Seong Y Lee, Brian C Joondeph, Steven S Houston, Kester Nahen, Nishant Mohan und 1 weitere

Ophthalmology Science · 2025-03 · 5 (2) · 100662

DOI: 10.1016/j.xops.2024.100662 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Ophthalmology Science, 2025-03. Verfasst von Eric W Schneider und 12 weiteren Autor:innen. Beteiligte Einrichtung: Tennessee Retina, PC, Nashville, Tennessee. Die Arbeit wurde bisher 19-mal zitiert (Crossref) und nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Notal Vision.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To validate the performance of the Notal OCT Analyzer (NOA) in processing self-administered OCT images from an OCT system designed for home use (home OCT [HOCT]) as part of a pivotal study aimed at achieving de novo United States Food and Drug Admininstration marketing authorization.

Design

A prospective quantitative cross-sectional artificial intelligence study.

Participants

The study enrolled adults aged ≥55 years diagnosed with neovascular age-related macular degeneration (nAMD) in ≥1 eligible eye with a best-corrected visual acuity of 20/320 or better.

Methods

Participants self-imaged 4 times on each of 2 HOCT devices and once with an in-office OCT (IO-OCT) device during a single visit. Scans were segmented by the NOA and human graders at a certified reading center (RC). Main Outcome Measures: Intradevice and interdevice repeatability and reproducibility of total retinal hyporeflective (TRO) volume estimation by the NOA on HOCT-acquired images as compared with that of RC graders on IO-OCT-acquired images. Additionally, to assess the performance of the NOA in segmentation of TRO over multiple B-scans as compared with RC graders.

Results

Self-imaging was performed successfully by 387 participants in 2451 of 2616 attempts (93.7%). For repeatability, the coefficient of variance for NOA was 11.1% for eyes with >10 volume units of TRO imaged with HOCT compared with 16.4% for RC graders segmenting IO-OCT images. The median DICE similarity coefficients for segmentation of TRO by NOA vs. Grader 1, Grader 2, and Grader 3; Grader 1 vs. Grader 2 and Grader 3; and Grader 2 vs. Grader 3 were 0.68, 0.68, 0.61, 0.72, 0.63, 0.70, respectively.

Conclusions

The performance of NOA supports the intended use of the system as a tool to monitor TRO at home between routine clinical visits in support of the management of nAMD. Financial Disclosures: Proprietary or commercial disclosure may be found in the Footnotes and Disclosures at the end of this article.

Schlagwörter

Stichwörter der Autor:innen

Age-related macular degenerationArtificial intelligenceHome OCTNeovascularTelemedicine

Autor:innen und Institutionen

  • Tennessee Retina, PC, Nashville, Tennessee.
  • Ophthalmic Consultants of Boston, Boston, Massachusetts.
  • Midwest Vision Research Foundation, Chesterfield, Missouri.
  • Retina Associates of Kentucky, Lexington, Kentucky.
  • Wagner Kapoor Institute, Norfolk, Virginia.
  • Georgia Retina, Marietta, Georgia.

Förderung

  • Notal Vision

Zitieren

APA

Schneider, E. W., Heier, J. S., Holekamp, N. M., Busquets, M. A., Wagner, A. L., Mukkamala, S. K., Riemann, C. D., Lee, S. Y., Joondeph, B. C., Houston, S. S., Nahen, K., Mohan, N., & Benyamini, G. (2025). Pivotal Trial Toward Effectiveness of Self-administered OCT in Neovascular Age-related Macular Degeneration. Report 2—Artificial Intelligence Analytics. Ophthalmology Science, 5(2), 100662. https://doi.org/10.1016/j.xops.2024.100662

Vancouver

Schneider EW, Heier JS, Holekamp NM, Busquets MA, Wagner AL, Mukkamala SK, et al. Pivotal Trial Toward Effectiveness of Self-administered OCT in Neovascular Age-related Macular Degeneration. Report 2—Artificial Intelligence Analytics. Ophthalmology Science. 2025;5(2):100662. doi:10.1016/j.xops.2024.100662

BibTeX
@article{Schneider2025,
  author = {Schneider, Eric W and Heier, Jeffrey S and Holekamp, Nancy M and Busquets, Miguel A and Wagner, Alan L and Mukkamala, S Krishna and Riemann, Christopher D and Lee, Seong Y and Joondeph, Brian C and Houston, Steven S and Nahen, Kester and Mohan, Nishant and Benyamini, Gidi},
  title = {Pivotal Trial Toward Effectiveness of Self-administered OCT in Neovascular Age-related Macular Degeneration. Report 2—Artificial Intelligence Analytics},
  journal = {Ophthalmology Science},
  year = {2025},
  volume = {5},
  number = {2},
  pages = {100662},
  doi = {10.1016/j.xops.2024.100662}
}