Circuit-dependent carbon dioxide rebreathing during continuous positive airway pressure

Ludovico Messineo, Bernard Hete, Ryan Diesem, William Noah

Sleep Medicine · 2024

DOI: 10.1016/j.sleep.2024.05.053 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sleep Medicine, 2024. Verfasst von Ludovico Messineo und 3 weiteren Autor:innen. Beteiligte Einrichtung: Division of Sleep and Circadian Disorders, Departments of Medicine and Neurology, Brigham & Women's Hospital & Harvard Medical School, Boston, MA, US…. Hauptthemen laut MEDLINE: Carbon Dioxide, Continuous Positive Airway Pressure, Sleep Apnea, Obstructive, Tidal Volume. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 23 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: The current standard treatment for obstructive sleep apnea (OSA), continuous positive airway pressure (CPAP), is characterized by a low adherence rate due to various factors including circuit-dependent carbon dioxide (CO2) rebreathing, which can exacerbated by disparate factors, such as low PAP, use of auto-titrating PAP or ramps. However, risk factors for rebreathing are often overlooked or poorly understood in clinical practice. Therefore, our objective was to evaluate the extent of rebreathing occurring with commonly used CPAP masks across varying PAPs, tidal volumes, and respiratory rates. METHODS: In a bench study, we assessed the rebreathing rate of nine masks interfacing a CPAP with a lung simulator providing different breathing respiratory rates (15 or 20 breaths/min) and tidal volumes (400, 500, 600, 700 and 750 mL). Additionally, a theoretical model was developed to describe the likelihood of CO2 rebreathing from four different masks at various breathing settings. RESULTS: Overall, all masks performed worse in situations characterized by low PAPs, high tidal volumes, and high respiratory rates. However, Dreamwear, Nuance, Siesta, Vitera, and particularly V2 masks exhibited greater susceptibility to rebreathing compared to F20, P10, Brevida, and Rio masks for the same variations of PAPs or ventilatory parameters. The mathematical model suggested that the risk of rebreathing for Rio, P10 and Nuance mask is negligible for respiratory rates of 10 breaths/min or below. CONCLUSIONS: Circuit-dependent CO2 rebreathing can be a common occurrence and warrants careful mask selection upon CPAP therapy initiation for optimal clinical outcomes.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Carbon DioxideContinuous Positive Airway PressureSleep Apnea, ObstructiveTidal VolumeMasksHumansRespiratory Rate

Stichwörter der Autor:innen

CPAP masksHypercapniaMask exhalation ports

Substanzen

Carbon Dioxide

Autor:innen und Institutionen

  • Division of Sleep and Circadian Disorders, Departments of Medicine and Neurology, Brigham & Women's Hospital & Harvard Medical School, Boston, MA, USA. Electronic address: ludovico.messineo@yahoo.it.
  • Sleep Centers of Middle Tennessee, USA.
  • Valley Inspired Products, USA.

Zitieren

APA

Messineo, L., Hete, B., Diesem, R., & Noah, W. (2024). Circuit-dependent carbon dioxide rebreathing during continuous positive airway pressure. Sleep Medicine. https://doi.org/10.1016/j.sleep.2024.05.053

Vancouver

Messineo L, Hete B, Diesem R, Noah W. Circuit-dependent carbon dioxide rebreathing during continuous positive airway pressure. Sleep Medicine. 2024. doi:10.1016/j.sleep.2024.05.053

BibTeX
@article{Messineo2024,
  author = {Messineo, Ludovico and Hete, Bernard and Diesem, Ryan and Noah, William},
  title = {Circuit-dependent carbon dioxide rebreathing during continuous positive airway pressure},
  journal = {Sleep Medicine},
  year = {2024},
  doi = {10.1016/j.sleep.2024.05.053}
}
Kostenlos starten →