Kurz gefasst

Fachartikel in der Zeitschrift SLAS Technology, 2026. Verfasst von Edy Mariano und 4 weiteren Autor:innen. Beteiligte Einrichtung: Swiss Cat+ West Hub, Ecole Polytechnique Fédérale de Lausanne (EPFL), Lausanne, Switzerland. Hauptthemen laut MEDLINE: Robotics, Laboratories, Automation, Laboratory, Safety. Die Arbeit nennt 45 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The emergence of autonomous laboratories is accelerating discovery in chemistry, drug discovery, materials science, and related fields by enabling high-throughput, data-driven experimentation. However, the integration of heterogeneous robotic systems, ranging from fixed manipulators to mobile platforms, introduces safety challenges that are not systematically addressed in newly established laboratories. In this context, this work aims to raise awareness of robotic safety among chemists and biologists leading laboratory automation projects who may have limited access to industrial robotics expertise. To support a preliminary evaluation of existing or newly developed automated laboratory systems, we explain and demonstrate the use of a simple, structured safety assessment methodology based on ISO standards and tailored to laboratory environments. The framework combines established robotics safety standards with laboratory-specific considerations, including chemical hazards, human-robot interaction, and dynamic workflows. To facilitate its adoption by scientists, the methodology is illustrated through a case study conducted at the Swiss CAT+ West Hub autonomous laboratory, focusing on a multi-instrument analytical platform integrating collaborative robotic arms and mobile robotic systems. The proposed framework follows a six-step iterative process encompassing system definition, hazard identification, scenario analysis, risk estimation, risk reduction, and validation. Its applicability was evaluated through the case study, in which sixteen hazards were identified, with robot-human collisions and chemical exposure representing the most critical risks. Experimental force and pressure measurements further demonstrated that widely used collaborative robots may exceed accepted safety thresholds under realistic operating conditions, particularly as a consequence of end-effector design and task-dependent motion characteristics. Risk mitigation strategies based on dynamic safety zoning, sensor-based human detection, and operational mode control were implemented to ensure compliance with safety requirements. The results highlight the need for systematic, context-specific safety assessments in autonomous laboratories and demonstrate that collaborative robots are not inherently safe without rigorous validation. This work provides a practical framework for the safe deployment of robotic systems in autonomous and digital laboratory environments.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

RoboticsLaboratoriesAutomation, LaboratorySafetyHumansAutonomous Robots

Stichwörter der Autor:innen

Autonomous laboratoriesCollaborative robotsCyber-physical systemsDigital chemical engineeringHuman–robot interactionLaboratory automationProcess safetyRisk assessmentRobotics safety

Autor:innen und Institutionen

  • Swiss Cat+ West Hub, Ecole Polytechnique Fédérale de Lausanne (EPFL), Lausanne, Switzerland.
  • School of Engineering and Management Vaud, HES SO University of Applied Sciences and Art Western Switzerland, Yverdon-les-bains, VD, Switzerland.
  • Swiss Cat+ West Hub, Ecole Polytechnique Fédérale de Lausanne (EPFL), Lausanne, Switzerland. Electronic address: pascal.mieville@epfl.ch.

Zitieren

APA

Mariano, E., Löwensberg, M., Bloesch, T., Costanzo, G. T., & Miéville, P. (2026). Robotic safety in self-driving laboratories. SLAS Technology. https://doi.org/10.1016/j.slast.2026.100463

Vancouver

Mariano E, Löwensberg M, Bloesch T, Costanzo GT, Miéville P. Robotic safety in self-driving laboratories. SLAS Technology. 2026. doi:10.1016/j.slast.2026.100463

BibTeX
@article{Mariano2026,
  author = {Mariano, Edy and Löwensberg, Maël and Bloesch, Théo and Costanzo, Giuseppe Tommaso and Miéville, Pascal},
  title = {Robotic safety in self-driving laboratories},
  journal = {SLAS Technology},
  year = {2026},
  doi = {10.1016/j.slast.2026.100463}
}
Kostenlos starten →