Long-term physicochemical and spectroscopic validation of a microfabricated Raman chip for biological reference samples
Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy · 2026-11 · 129118
DOI: 10.1016/j.saa.2026.129118 ↗Kurz gefasst
Fachartikel in der Zeitschrift Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy, 2026-11. Verfasst von Markus Salbreiter und 7 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Physical Chemistry and Abbe Center of Photonics, Friedrich Schiller University, Jena 07743, Germany; InfectoGnostics Research Campus Jen…. Die Arbeit nennt 71 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: European Regional Development Fund, European Cooperation in Science and Technology, Bundesministerium für Forschung Technologie und Raumfahrt Informationstechnikzentrum Bund.
Abstract (Originalsprache)
The lack of biologically relevant reference materials remains a major challenge for the standardization of Raman spectroscopy. Here, we systematically evaluate the long-term physicochemical and spectroscopic stability of a previously established microfabricated aluminium-coated silicon Raman chip for biological reference applications. The platform combines an aluminium-coated Raman-compatible measurement surface with a microscope slide-compatible format for integration into conventional Raman microscopy workflows. Longitudinal X-ray photoelectron spectroscopy revealed comparatively small changes in the surface composition of individual aluminium-coated chips stored at room temperature or under refrigerated conditions over six months, whereas more pronounced surface modifications were observed for the chip investigated under humid storage conditions. Using bacterial endospores as a model biological system, longitudinal Raman measurements at 532 and 785 nm excitation showed preservation of characteristic Raman fingerprints without systematic spectral drift or substrate-derived interference. While increasing fluorescence increasingly obscured the Raman signals at 532 nm after approximately four weeks, characteristic Raman spectra remained detectable at 785 nm throughout the twelve-week monitoring period. These findings provide initial evidence of temporal physicochemical and spectroscopic stability within the investigated chip preparations and support further validation of the platform for biological Raman reference applications.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Institute of Physical Chemistry and Abbe Center of Photonics, Friedrich Schiller University, Jena 07743, Germany; InfectoGnostics Research Campus Jena, Center of Applied Research, 07743 Jena, Germany.
- Institute of Physical Chemistry and Abbe Center of Photonics, Friedrich Schiller University, Jena 07743, Germany.
- InfectoGnostics Research Campus Jena, Center of Applied Research, 07743 Jena, Germany; Leibniz Institute of Photonic Technology, Member of the Leibniz Research Alliance - Leibniz Health Technologies, 07745 Jena, Germany.
- Institute of Physical Chemistry and Abbe Center of Photonics, Friedrich Schiller University, Jena 07743, Germany; Jena Center for Soft Matter (JCSM), Philosophenweg 7, 07743 Jena, Germany.
- Institute of Physical Chemistry and Abbe Center of Photonics, Friedrich Schiller University, Jena 07743, Germany; InfectoGnostics Research Campus Jena, Center of Applied Research, 07743 Jena, Germany; Leibniz Institute…
- Institute of Physical Chemistry and Abbe Center of Photonics, Friedrich Schiller University, Jena 07743, Germany; Leibniz Institute of Photonic Technology, Member of the Leibniz Research Alliance - Leibniz Health Techno…
Förderung
- European Regional Development Fund
- European Cooperation in Science and Technology
- Bundesministerium für Forschung Technologie und Raumfahrt Informationstechnikzentrum Bund
- German Research Foundation
Zitieren
APA
Salbreiter, M., Neumann, C., Wagenhaus, A., Pahlow, S., Silge, A., Turchanin, A., Popp, J., & Bocklitz, T. W. (2026). Long-term physicochemical and spectroscopic validation of a microfabricated Raman chip for biological reference samples. Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy, 129118. https://doi.org/10.1016/j.saa.2026.129118
Vancouver
Salbreiter M, Neumann C, Wagenhaus A, Pahlow S, Silge A, Turchanin A, et al. Long-term physicochemical and spectroscopic validation of a microfabricated Raman chip for biological reference samples. Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy. 2026:129118. doi:10.1016/j.saa.2026.129118
BibTeX
@article{Salbreiter2026,
author = {Salbreiter, Markus and Neumann, Christof and Wagenhaus, Annette and Pahlow, Susanne and Silge, Anja and Turchanin, Andrey and Popp, Jürgen and Bocklitz, Thomas W},
title = {Long-term physicochemical and spectroscopic validation of a microfabricated Raman chip for biological reference samples},
journal = {Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy},
year = {2026},
pages = {129118},
doi = {10.1016/j.saa.2026.129118}
}