Terahertz metamaterial sensor-based detection of procymidone and chlorpyrifos residues in Ligusticum chuanxiong
Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy · 2026
DOI: 10.1016/j.saa.2026.127652 ↗Kurz gefasst
Fachartikel in der Zeitschrift Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy, 2026. Verfasst von Yuting Shao und 6 weiteren Autor:innen. Beteiligte Einrichtung: Terahertz Technology Innovation Research Institute, Terahertz Spectrum and Imaging Technology Cooperative Innovation Center, Shanghai Key Lab of Mode…. Hauptthemen laut MEDLINE: Ligusticum, Chlorpyrifos, Pesticide Residues, Terahertz Spectroscopy. Die Arbeit nennt 46 Literaturangaben. Gefördert von: National Natural Science Foundation of China.
Abstract (Originalsprache)
Pesticide residues in Traditional Chinese Medicine (TCM) have raised significant concerns due to their potential health risks. Conventional pesticide residues detection methods are often limited by prolonged detection times and complex sample pretreatment procedures. Therefore, developing a rapid and highly sensitive method for pesticide residue detection is essential. In this paper, we present a terahertz (THz) metamaterial sensor for rapid detection of trace procymidone and chlorpyrifos residues in Ligusticum chuanxiong. The sensor exhibits a theoretical sensitivity of 360 GHz/RIU and offers a minimum detectable level of 1 μg/L, while the entire workflow can be completed within 30 min. Based on the THz metamaterial sensor, quantitative concentration-frequency shift models for chlorpyrifos and procymidone in organic Ligusticum chuanxiong were established (fitting coefficient R2 > 0.98). These models, combined with the proposed Leave-One-Out Cross-Validation (LOO-CV) based calibration method, were applied to the detection of pesticide residues in commercial Ligusticum chuanxiong. The final average prediction error was approximately 3.8%, providing a novel and rapid testing solution for the quality control of TCM.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Terahertz Technology Innovation Research Institute, Terahertz Spectrum and Imaging Technology Cooperative Innovation Center, Shanghai Key Lab of Modern Optical System, University of Shanghai for Science and Technology,…
- School of Artificial Intelligence in Traditional Chinese Medicine, Shanghai 201203, China.
- School of Artificial Intelligence in Traditional Chinese Medicine, Shanghai 201203, China. Electronic address: vincent.tang@shutcm.edu.cn.
Förderung
- National Natural Science Foundation of China
Zitieren
APA
Shao, Y., Zhu, X., Liu, B., Peng, Y., Zhu, Y., Tang, W., & Shao, Y. (2026). Terahertz metamaterial sensor-based detection of procymidone and chlorpyrifos residues in Ligusticum chuanxiong. Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy. https://doi.org/10.1016/j.saa.2026.127652
Vancouver
Shao Y, Zhu X, Liu B, Peng Y, Zhu Y, Tang W, et al. Terahertz metamaterial sensor-based detection of procymidone and chlorpyrifos residues in Ligusticum chuanxiong. Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy. 2026. doi:10.1016/j.saa.2026.127652
BibTeX
@article{Shao2026,
author = {Shao, Yuting and Zhu, Xinyu and Liu, Bingwei and Peng, Yan and Zhu, Yiming and Tang, Wenchao and Shao, Yongni},
title = {Terahertz metamaterial sensor-based detection of procymidone and chlorpyrifos residues in Ligusticum chuanxiong},
journal = {Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy},
year = {2026},
doi = {10.1016/j.saa.2026.127652}
}