An autocatalytic hybridization circuit-based FRET aptasensor for detection of low-abundant sulfameter in human serum
Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy · 2025
DOI: 10.1016/j.saa.2024.125241 ↗Kurz gefasst
Fachartikel in der Zeitschrift Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy, 2025. Verfasst von Yanping He und 4 weiteren Autor:innen. Beteiligte Einrichtung: Laboratory of Functionalized Molecular Solids, Ministry of Education, Anhui Province Key Laboratory of Biomedical Materials and Chemical Measurement,…. Hauptthemen laut MEDLINE: Fluorescence Resonance Energy Transfer, Aptamers, Nucleotide, Nucleic Acid Hybridization, Biosensing Techniques. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 44 Literaturangaben. Gefördert von: Anhui Normal University.
Abstract (Originalsprache)
Exposure to antibiotics is considered a potential risk factor for human health. Yet, the extensive and cost-effective detection of low-abundant antibiotics in complex matrices remains a significant challenge. Herein, an aptamer and an autocatalytic hybridization circuit (AHC) were used to fabricate a fluorescence resonance energy transfer (FRET) platform to detect sulfameter (SME) in human serum. The AHC system comprised two mutually motivated hybridization chain reactions (HCR) modules, ultimately producing long-branched DNA copolymeric nanowires. This mutually reciprocal activation of two HCR modules enables continuous signal amplification, providing the AHC system with wide linear range and high sensitivity for the SME detection. Compared to the HCR-based aptasensor, the AHC-based aptasensor exhibited a wider linear range and improved sensitivity (3.3 times greater). Under optimal conditions, the fluorescent AHC-based aptasensor demonstrated a linear range (R2 was 0.996) from 0.5 to 2000 nM, with a low detection limit of 0.301 nM (S/N = 3). The fluorescent aptasensor was also validated by SME-spiked human serum samples, showing average recoveries ranging from 96.40 % to 109.30 %, with a relative standard deviation below 10.45 %. Furthermore, when tested on six human serum samples, the aptasensor results were consistent with those obtained from the commercial ELISA method. These findings demonstrate that the proposed aptasensor provides a promising approach for the practical monitoring of low-abundant SME in human serum.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Laboratory of Functionalized Molecular Solids, Ministry of Education, Anhui Province Key Laboratory of Biomedical Materials and Chemical Measurement, College of Chemistry and Materials Science, Anhui Normal University,…
Förderung
- Anhui Normal University
Zitieren
APA
He, Y., Tian, R., Shen, W., Zhang, J., & Tao, C. (2025). An autocatalytic hybridization circuit-based FRET aptasensor for detection of low-abundant sulfameter in human serum. Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy. https://doi.org/10.1016/j.saa.2024.125241
Vancouver
He Y, Tian R, Shen W, Zhang J, Tao C. An autocatalytic hybridization circuit-based FRET aptasensor for detection of low-abundant sulfameter in human serum. Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy. 2025. doi:10.1016/j.saa.2024.125241
BibTeX
@article{He2025,
author = {He, Yanping and Tian, Ruifen and Shen, Weili and Zhang, Jingrui and Tao, Chen},
title = {An autocatalytic hybridization circuit-based FRET aptasensor for detection of low-abundant sulfameter in human serum},
journal = {Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy},
year = {2025},
doi = {10.1016/j.saa.2024.125241}
}
EUVIMED