Effect of a calibration aid and the intraoral scanner on the registration of a partially edentulous maxilla: An in vitro study
The Journal of Prosthetic Dentistry · 2025-02 · 133 (2) · 560-567
DOI: 10.1016/j.prosdent.2024.02.019 ↗Kurz gefasst
Fachartikel in der Zeitschrift The Journal of Prosthetic Dentistry, 2025-02. Verfasst von Sophia Boehm und 4 weiteren Autor:innen. Beteiligte Einrichtung: Assistant Professor, Department of Prosthetic Dentistry, Heidelberg University Hospital, University of Heidelberg, Heidelberg, Germany. Hauptthemen laut MEDLINE: Maxilla, Jaw, Edentulous, Partially. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Medizinischen Fakultät Heidelberg, Universität Heidelberg.
Abstract (Originalsprache)
STATEMENT OF PROBLEM: Precise data are a prerequisite for accurately fitting restorations. Therefore, intraoral scanners have to be reliable. Data on differences between the same model of intraoral scanning systems are lacking. PURPOSE: This in vitro study evaluated differences in the scanning accuracy of a partially edentulous maxilla between combinations of new intraoral scanners of a single model from the same manufacturer (Primescan; Dentsply Sirona) and several calibration aids. MATERIAL AND METHODS: The in vitro reference model simulated a maxilla with 6 prepared teeth for a complete arch fixed partial denture. Five precision ceramic balls were used to detect dimensional deviation between the reference model and the scan. Distances were divided into 4 categories, from short distances between 2 neighboring precision balls to the cross-arch distance with the scan path comprising all 5 balls. For each combination of 4 new intraoral scanners and their respective calibration aids, 12 model registrations were generated. The data were statistically analyzed using ANOVA (α=.05). RESULTS: Distance deviations increased with increasing scan path length and were significantly affected by the covariates "scanner" (P≤.023) and, for 3 of 4 distance categories, the "calibration aid" (short, medium, and long distances: P≤.013). For short and medium distances, acceptable scanning results were achieved for all test groups. The largest deviation was 539 µm for the cross-arch distance. CONCLUSIONS: Scanning accuracy depends on the scanner and the calibration aid used, in particular, for spans exceeding a single quadrant.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Assistant Professor, Department of Prosthetic Dentistry, Heidelberg University Hospital, University of Heidelberg, Heidelberg, Germany.
- Assistant Professor, Institute of Medical Biometry, Heidelberg University Hospital, University of Heidelberg, Heidelberg, Germany.
- Full Professor and Director of Department, Department of Prosthetic Dentistry, Heidelberg University Hospital, University of Heidelberg, Heidelberg, Germany.
- Assistant Professor, Department of Prosthetic Dentistry, Heidelberg University Hospital, University of Heidelberg, Heidelberg, Germany. Electronic address: moritz.waldecker@med.uni-heidelberg.de.
Förderung
- Medizinischen Fakultät Heidelberg, Universität Heidelberg
Zitieren
APA
Boehm, S., Rues, S., Balzer, A., Rammelsberg, P., & Waldecker, M. (2025). Effect of a calibration aid and the intraoral scanner on the registration of a partially edentulous maxilla: An in vitro study. The Journal of Prosthetic Dentistry, 133(2), 560-567. https://doi.org/10.1016/j.prosdent.2024.02.019
Vancouver
Boehm S, Rues S, Balzer A, Rammelsberg P, Waldecker M. Effect of a calibration aid and the intraoral scanner on the registration of a partially edentulous maxilla: An in vitro study. The Journal of Prosthetic Dentistry. 2025;133(2):560-567. doi:10.1016/j.prosdent.2024.02.019
BibTeX
@article{Boehm2025,
author = {Boehm, Sophia and Rues, Stefan and Balzer, Alexandra and Rammelsberg, Peter and Waldecker, Moritz},
title = {Effect of a calibration aid and the intraoral scanner on the registration of a partially edentulous maxilla: An in vitro study},
journal = {The Journal of Prosthetic Dentistry},
year = {2025},
volume = {133},
number = {2},
pages = {560--567},
doi = {10.1016/j.prosdent.2024.02.019}
}
EUVIMED