Vergleichsstudie

Scanning accuracy and scanning area discrepancies of intraoral digital scans acquired at varying scanning distances and angulations among 4 different intraoral scanners

Heather Button, John C Kois, Abdul B Barmak, Jonathan M Zeitler, Vygandas Rutkunas, Marta Revilla-León

The Journal of Prosthetic Dentistry · 2024-11 · 132 (5) · 1044-1060

DOI: 10.1016/j.prosdent.2023.01.025 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift The Journal of Prosthetic Dentistry, 2024-11. Verfasst von Heather Button und 5 weiteren Autor:innen. Beteiligte Einrichtung: Clinical Manager, Kois Center, Seattle, Wash. Hauptthemen laut MEDLINE: Computer-Aided Design. Die Arbeit wurde bisher 71-mal zitiert (Crossref) und nennt 27 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

STATEMENT OF PROBLEM: The accuracy of intraoral scanners (IOSs) can be affected by operator handling; however, the scanning area and accuracy discrepancies acquired at different scanning distances and angulations among IOSs remain uncertain. PURPOSE: The objective of this in vitro study was to compare the scanning area and scanning accuracy of the intraoral digital scans obtained at 3 scanning distances with 4 different scanning angulations among 4 different IOSs. MATERIAL AND METHODS: A reference device (reference file) was designed with 4 inclinations (0, 15, 30, and 45 degrees) and printed. Four groups were created based on the IOS: i700, TRIOS4, CS 3800, and iTero scanners. Four subgroups were generated depending on the scanning angulation (0, 15, 30, and 45 degrees). Each subgroup was divided into 3 subgroups based on the scanning distance: 0, 2, and 4 mm (N=720, n=15). The reference devices were positioned in a z-axis calibrated platform for standardizing the scanning distance. In the i700-0-0 subgroup, the 0-degree reference device was positioned in the calibrated platform. The wand of the IOS was positioned in a supporting framework with a 0-mm scanning distance, and the scans were acquired. In the i700-0-2 subgroup, the platform was lowered for a 2-mm scanning distance followed by the specimen acquisition. In the i700-0-4 subgroup, the platform was further lowered for a 4-mm scanning distance, and the scans were obtained. For the i700-15, i700-30, and i700-45 subgroups, the same procedures were carried out as in the i700-0 subgroups respectively, but with the 10-, 15-, 30-, or 45-degree reference device. Similarly, the same procedures were completed for all the groups with the corresponding IOS. The area of each scan was measured. The reference file was used to measure the discrepancy with the experimental scans by using the root mean square (RMS) error. Three-way ANOVA and post hoc Tukey pairwise comparison tests were used to analyze the scanning area data. Kruskal-Wallis and multiple pairwise comparison tests were used to analyze the RMS data (α=.05). RESULTS: IOS (P .999). All the scanning distance groups were different from each other (P

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Computer-Aided DesignHumansDental Impression TechniqueImage Processing, Computer-AssistedImaging, Three-Dimensional

Autor:innen und Institutionen

  • Clinical Manager, Kois Center, Seattle, Wash.
  • Founder and Director Kois Center, Seattle, Wash; Affiliate Professor, Graduate Prosthodontics, Department of Restorative Dentistry, University of Washington, Seattle, Wash; and Private Practice, Seattle, Wash.
  • Assistant Professor, Clinical Research and Biostatistics, Eastman Institute of Oral Health, University of Rochester Medical Center, Rochester, NY.
  • Director of IT, Kois Center, Seattle, Wash.
  • Director Digitorum Research Center, Vilnius, Lithuania; and Professor, Department of Prosthodontics, Institute of Odontology, Faculty of Medicine, Vilnius University, Vilnius, Lithuania.
  • Affiliate Assistant Professor, Graduate Prosthodontics, Department of Restorative Dentistry, School of Dentistry, University of Washington, Seattle, Wash; Faculty and Director of Research and Digital Dentistry, Kois Cen…

Zitieren

APA

Button, H., Kois, J. C., Barmak, A. B., Zeitler, J. M., Rutkunas, V., & Revilla-León, M. (2024). Scanning accuracy and scanning area discrepancies of intraoral digital scans acquired at varying scanning distances and angulations among 4 different intraoral scanners. The Journal of Prosthetic Dentistry, 132(5), 1044-1060. https://doi.org/10.1016/j.prosdent.2023.01.025

Vancouver

Button H, Kois JC, Barmak AB, Zeitler JM, Rutkunas V, Revilla-León M. Scanning accuracy and scanning area discrepancies of intraoral digital scans acquired at varying scanning distances and angulations among 4 different intraoral scanners. The Journal of Prosthetic Dentistry. 2024;132(5):1044-1060. doi:10.1016/j.prosdent.2023.01.025

BibTeX
@article{Button2024,
  author = {Button, Heather and Kois, John C and Barmak, Abdul B and Zeitler, Jonathan M and Rutkunas, Vygandas and Revilla-León, Marta},
  title = {Scanning accuracy and scanning area discrepancies of intraoral digital scans acquired at varying scanning distances and angulations among 4 different intraoral scanners},
  journal = {The Journal of Prosthetic Dentistry},
  year = {2024},
  volume = {132},
  number = {5},
  pages = {1044--1060},
  doi = {10.1016/j.prosdent.2023.01.025}
}
Kostenlos starten →