EUVIMEDEUVIMED entdecken
Crossref

Mechanical complications of implant-supported restorations with internal conical connection implants: A 14-year retrospective study

Yuseung Yi, Seong-Joo Heo, Jai-Young Koak, Seong-Kyun Kim

The Journal of Prosthetic Dentistry · 2023-05 · 129 (5) · 732-740

DOI: 10.1016/j.prosdent.2021.06.053 ↗

Kurz gefasst

Fachartikel in der Zeitschrift The Journal of Prosthetic Dentistry, 2023-05. Verfasst von Yuseung Yi und 3 weiteren Autor:innen. Beteiligte Einrichtung: Clinical Professor, Department of Prosthodontics, Seoul National University Dental Hospital, Seoul, Republic of Korea. Hauptthemen laut MEDLINE: Dental Implants. Die Arbeit wurde bisher 28-mal zitiert (Crossref) und nennt 34 Literaturangaben. Gefördert von: Ministry of Food and Drug Safety, Ministry of Science, ICT and Future Planning, Ministry of Trade, Industry and Energy.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

STATEMENT OF PROBLEM: Internal conical connections have become the primary choice for implant-supported restorations. However, studies that identified the risk indicators for mechanical complications and diagnosed the prognosis of the implant-supported restorations are lacking. PURPOSE: The purpose of this retrospective clinical study was to evaluate the incidence and consequences of mechanical complications in components of internal conical connection implants and to analyze the risk indicators. MATERIAL AND METHODS: A total of 428 patients with 898 internal conical connection implants were included in the study, and mechanical complications over the 14-year observation period were evaluated. The Cox proportional hazard regression model was used to analyze significant effects on mechanical complications, which were presented as hazard ratio and 95% confidence interval. RESULTS: Mechanical complications occurred in 430 (47.9%) implants and 248 (57.9%) patients: screw loosening (46.4% of implants, 56.8% of patients); screw fracture (2.6% of implants, 4.4% of patients); abutment fracture (11.4% of implants, 21.3% of patients); and implant fracture (3.5% of implants, 5.4% of patients). Implant restorations replacing molars showed the highest risk for mechanical complication (hazard ratio 12.82; 95% confidence interval 2.73-60.31) and for fracture of all components. Men had a higher risk of mechanical complication than women (hazard ratio 2.00; 95% confidence interval 1.55-2.59), and the risk of fracture was higher in all components. With increased splinted implants, the risk of mechanical complication (hazard ratio 0.67; 95% confidence interval 0.49-0.93) and component fracture (hazard ratio 0.73; 95% confidence interval 0.29-0.89) decreased. Gold screws had a lower risk of screw loosening (hazard ratio 0.74; 95% confidence interval 0.58-0.94) than titanium screws, but a higher risk of fracture (hazard ratio 3.45; 95% confidence interval 1.42-8.36). The smaller the implant diameter, the higher the risk of implant fracture (hazard ratio 0.01; 95% confidence interval 0.00-0.05). CONCLUSIONS: Abutments were most frequently fractured among the components of internal conical connection type implants. Molar implant-supported restorations and male patients had higher risks of mechanical complications, and as the number of splinted implants in a prosthesis increased, the risk decreased. Gold screws showed less risk of screw loosening and higher risk of fracture than titanium screws. Narrow-diameter implants had a higher risk of fracture.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Dental ImplantsHumansMaleFemaleRetrospective StudiesTitaniumDental Prosthesis, Implant-SupportedDental AbutmentsDental Restoration FailureGoldDental Implant-Abutment Design

Substanzen

Dental ImplantsTitaniumGold

Autor:innen und Institutionen

  • Clinical Professor, Department of Prosthodontics, Seoul National University Dental Hospital, Seoul, Republic of Korea.
  • Professor, Department of Prosthodontics & Dental Research Institute, School of Dentistry, Seoul National University, Seoul, Republic of Korea.
  • Professor, Department of Prosthodontics & Dental Research Institute, School of Dentistry, Seoul National University, Seoul, Republic of Korea. Electronic address: 0504heo@hanmail.net.

Förderung

  • Ministry of Food and Drug Safety
  • Ministry of Science, ICT and Future Planning
  • Ministry of Trade, Industry and Energy
  • Ministry of Health and Welfare

Zitieren

APA

Yi, Y., Heo, S. J., Koak, J. Y., & Kim, S. K. (2023). Mechanical complications of implant-supported restorations with internal conical connection implants: A 14-year retrospective study. The Journal of Prosthetic Dentistry, 129(5), 732-740. https://doi.org/10.1016/j.prosdent.2021.06.053

Vancouver

Yi Y, Heo SJ, Koak JY, Kim SK. Mechanical complications of implant-supported restorations with internal conical connection implants: A 14-year retrospective study. The Journal of Prosthetic Dentistry. 2023;129(5):732-740. doi:10.1016/j.prosdent.2021.06.053

BibTeX
@article{Yi2023,
  author = {Yi, Yuseung and Heo, Seong-Joo and Koak, Jai-Young and Kim, Seong-Kyun},
  title = {Mechanical complications of implant-supported restorations with internal conical connection implants: A 14-year retrospective study},
  journal = {The Journal of Prosthetic Dentistry},
  year = {2023},
  volume = {129},
  number = {5},
  pages = {732--740},
  doi = {10.1016/j.prosdent.2021.06.053}
}