Validierungsstudie in der Zeitschrift Placenta, 2024-12. Verfasst von Farah Deeba und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electrical and Computer Engineering, University of British Columbia, Vancouver, British Columbia, Canada; Department of Electrical and …. Hauptthemen laut MEDLINE: Placenta, Pre-Eclampsia, Ultrasonography, Prenatal. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Canadian Institutes of Health Research, Natural Sciences and Engineering Research Council of Canada, Schlumberger Foundation.
Abstract (Originalsprache)
INTRODUCTION: Placenta-mediated diseases are associated with structural changes in the placenta. Quantitative Ultrasound (QUS) imaging measures the acoustic properties of the tissue, which are correlated to the underlying tissue structure. We aimed to develop and validate a diagnostic prediction model using QUS measurements for pre-eclampsia (PE) and small-for-gestational-age (SGA) fetuses/neonates. METHODS: For this prospective case-control study, placentas were collected from a group of women who delivered via cesarean section at BC Women's Hospital, Vancouver, Canada. Ultrasound data were collected and processed to compute three QUS parameters, namely, attenuation coefficient estimate (ACE), integrated backscatter coefficient (IBC), and effective scatterer diameter (ESD) from the placentas. We developed a logistic regression model using QUS parameters as predictors. The primary outcome was the occurrence of SGA and PE. RESULTS: The dataset consisted of 47 placentas, of which 25 placentas were complicated by SGA/PE. The final placenta-QUS model included quadratic and interaction terms of ACE, IBC, and ESD parameters. The placenta-QUS model was well-calibrated, with a calibration slope of 0.99 (0.57-1.05) and a calibration intercept of 0.003 (-0.02 - 0.22). The model predicted the SGA/PE complicated pregnancies with an apparent Area Under the Receive Operating Characteristic Curve (AUROC) of 0.89 (95 % CI: 0.78-0.98). The optimism-adjusted AUROC was 0.88 (95 % CI: 0.78-0.98). DISCUSSION: A model for SGA and PE has been developed using QUS measures from the placenta ex vivo. The model showed promising performance in detecting SGA/PE. Future studies will be performed to assess the model performance using QUS measures in utero.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Electrical and Computer Engineering, University of British Columbia, Vancouver, British Columbia, Canada; Department of Electrical and Computer Engineering, University of North Carolina at Charlotte, Charl…
- Department of Electrical and Computer Engineering, University of British Columbia, Vancouver, British Columbia, Canada.
- Department of Ultrasound, BC Women's Hospital, Vancouver, British Columbia, Canada.
- Department of Pathology and Laboratory Medicine, University of British Columbia, British Columbia, Canada.
- Department of Radiology, University of British Columbia, Vancouver, British Columbia, Canada.
- Department of Obstetrics and Gynecology, University of British Columbia, Vancouver, British Columbia, Canada.
Förderung
- Canadian Institutes of Health Research
- Natural Sciences and Engineering Research Council of Canada
- Schlumberger Foundation
- Microsoft Research
- Microsoft
Zitieren
APA
Deeba, F., Hu, R., Lessoway, V., Terry, J., Pugash, D., Mayer, C., Hutcheon, J., & Rohling, R. (2024). Development and validation of the placenta-QUS model for the detection of placenta-mediated diseases using quantitative ultrasound measurements: An Ex Vivo proof-of-concept study. Placenta, 158, 293-300. https://doi.org/10.1016/j.placenta.2024.11.004
Vancouver
Deeba F, Hu R, Lessoway V, Terry J, Pugash D, Mayer C, et al. Development and validation of the placenta-QUS model for the detection of placenta-mediated diseases using quantitative ultrasound measurements: An Ex Vivo proof-of-concept study. Placenta. 2024;158:293-300. doi:10.1016/j.placenta.2024.11.004
BibTeX
@article{Deeba2024,
author = {Deeba, Farah and Hu, Ricky and Lessoway, Victoria and Terry, Jefferson and Pugash, Denise and Mayer, Chantal and Hutcheon, Jennifer and Rohling, Robert},
title = {Development and validation of the placenta-QUS model for the detection of placenta-mediated diseases using quantitative ultrasound measurements: An Ex Vivo proof-of-concept study},
journal = {Placenta},
year = {2024},
volume = {158},
pages = {293--300},
doi = {10.1016/j.placenta.2024.11.004}
}
EUVIMED