Analytical validation of a MALDI-TOF mass spectrometry method for serum IGF-1 measurement and its comparison with LC-MS/MS and CLIA

Ruiquan Cao, Huan Ding, Yong Xia, Xueting Ma, Anping Xu, Hou-Long Luo, Ling Ji

Practical Laboratory Medicine · 2026-12 · 52 · e00559

DOI: 10.1016/j.plabm.2026.e00559 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Practical Laboratory Medicine, 2026-12. Verfasst von Ruiquan Cao und 6 weiteren Autor:innen. Beteiligte Einrichtung: School of Medical Technology, Guangdong Medical University, Zhanjiang, Guangdong, China. Die Arbeit nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Department of Science and Technology of Guangdong Province.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Insulin-like growth factor 1 (IGF-1) is an important biomarker in the diagnosis and monitoring of growth hormone-related disorders. MALDI-TOF mass spectrometry has potential for high-throughput IGF-1 measurement, but its analytical performance for routine clinical application requires systematic validation.

Methods

Analytical performance of a MALDI-TOF MS-based IGF-1 assay was evaluated following CLSI guidelines, including imprecision, accuracy, linearity, limit of quantitation (LOQ), analytical specificity, sample stability, and sample type comparison. Method comparison was performed against LC-MS/MS and CLIA using 287 patient serum samples.

Results

Intra-assay coefficients of variation (CVs) were 3.66% and 1.82% at low (26.52 ng/mL) and high (375.49 ng/mL) concentrations, respectively; inter-assay CVs were 6.21% and 3.21%. The assay was linear (r = 0.998) with an LOQ of 5 ng/mL. Recoveries ranged from 87.70% to 91.73%. Cross-reactivity with IGF-2 and insulin was negligible. Serum IGF-1 remained stable for 3 days at 25°C, 9 days at 2-8°C, and 12 weeks at -20°C, and tolerated five freeze-thaw cycles. EDTA plasma and serum showed acceptable agreement (mean bias 0.36%-6.62%) for interchangeable use. Method comparison showed strong correlations: MALDI-TOF MS vs. LC-MS/MS (Passing-Bablok: Y = -5.602 + 0.746X, r = 0.996) and vs. CLIA (Y = 2.378 + 0.739X, r = 0.994).

Conclusions

The MALDI-TOF MS method achieved acceptable analytical performance for serum IGF-1 measurement, including precision, linearity, stability, and specificity. The observed systematic bias relative to LC-MS/MS and CLIA suggests that method-specific reference intervals are needed. MALDI-TOF MS is a reliable and high-throughput alternative for clinical IGF-1 testing.

Schlagwörter

Stichwörter der Autor:innen

Analytical performanceChemiluminescence immunoassayInsulin-like growth factor 1Liquid chromatography-tandem mass spectrometryMALDI-TOF mass spectrometry

Autor:innen und Institutionen

  • School of Medical Technology, Guangdong Medical University, Zhanjiang, Guangdong, China.
  • Department of Laboratory Medicine, Peking University Shenzhen Hospital, Shenzhen, Guangdong, China.
  • Intelligene Biosystems Co., Ltd., Qingdao, Shangdong, China.

Förderung

  • Department of Science and Technology of Guangdong Province

Zitieren

APA

Cao, R., Ding, H., Xia, Y., Ma, X., Xu, A., Luo, H. L., & Ji, L. (2026). Analytical validation of a MALDI-TOF mass spectrometry method for serum IGF-1 measurement and its comparison with LC-MS/MS and CLIA. Practical Laboratory Medicine, 52, e00559. https://doi.org/10.1016/j.plabm.2026.e00559

Vancouver

Cao R, Ding H, Xia Y, Ma X, Xu A, Luo HL, et al. Analytical validation of a MALDI-TOF mass spectrometry method for serum IGF-1 measurement and its comparison with LC-MS/MS and CLIA. Practical Laboratory Medicine. 2026;52:e00559. doi:10.1016/j.plabm.2026.e00559

BibTeX
@article{Cao2026,
  author = {Cao, Ruiquan and Ding, Huan and Xia, Yong and Ma, Xueting and Xu, Anping and Luo, Hou-Long and Ji, Ling},
  title = {Analytical validation of a MALDI-TOF mass spectrometry method for serum IGF-1 measurement and its comparison with LC-MS/MS and CLIA},
  journal = {Practical Laboratory Medicine},
  year = {2026},
  volume = {52},
  pages = {e00559},
  doi = {10.1016/j.plabm.2026.e00559}
}
Kostenlos starten →