The „EPiQ“-Study (Evaluation of preanalytical quality): S-Monovette® in manual aspiration mode drastically reduces hemolytic samples in head-to-head study
Practical Laboratory Medicine · 2021-11 · 27 · e00252
DOI: 10.1016/j.plabm.2021.e00252 ↗Kurz gefasst
Fachartikel in der Zeitschrift Practical Laboratory Medicine, 2021-11. Verfasst von Laura Millius und 6 weiteren Autor:innen. Beteiligte Einrichtung: HFR Fribourg, Hôpital Cantonal, Fribourg, FR, Switzerland. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
BACKGROUND: Hemolytic blood samples are the number one cause for specimen rejection at emergency departments. Triggered by unsuitable blood sampling material or incorrect handling and a related strong vacuum force, hemolytic samples often must be retaken. The objective of this study was to assess whether correct manual aspiration using S-Monovette® could reduce the number of hemolytic samples. METHODS: Between January and April 2019, a head-to-head study was conducted. Whereas in the first eight weeks, all specimens were collected using Vacutainer®, in the second eight weeks, blood was taken using S-Monovette® in aspiration mode. Specimens were categorized into five classes (0-30, 31-50, 51-75, 76-100, and 101+ mg/dl of cell-free hemoglobin) and for the statistical analyses, all samples exceeding 30 mg/dl were classified as hemolytic. RESULTS: Data were collected on 4794 blood specimens (Vacutainer®: 2634 samples, S-Monovette®: 2160 samples). While the percentage of non-hemolytic samples (HI of 0-30 mg/dl) was substantially higher for specimens drawn by S-Monovette® (95.7 %) than Vacutainer® (83.0 %), the opposite was true for all HI categories above 30 mg/dl. Importantly, the reduction of hemolytic samples took place immediately following the imposition of S-Monovette® and remained stable at a low level until the end of the study. CONCLUSIONS: Based on our results, we conclude that switching to S-Monovette® in manual aspiration mode in the blood sampling process could be highly beneficial, not only from a financial point of view, but also with regards to reducing unnecessary tasks and stress for nursing staff and improving patient outcome overall.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- HFR Fribourg, Hôpital Cantonal, Fribourg, FR, Switzerland.
Zitieren
APA
Millius, L., Riedo, E., Caron, T., Belissent, J., Fellay, B., Ribordy, V., & Magnin, J. L. (2021). The „EPiQ“-Study (Evaluation of preanalytical quality): S-Monovette® in manual aspiration mode drastically reduces hemolytic samples in head-to-head study. Practical Laboratory Medicine, 27, e00252. https://doi.org/10.1016/j.plabm.2021.e00252
Vancouver
Millius L, Riedo E, Caron T, Belissent J, Fellay B, Ribordy V, et al. The „EPiQ“-Study (Evaluation of preanalytical quality): S-Monovette® in manual aspiration mode drastically reduces hemolytic samples in head-to-head study. Practical Laboratory Medicine. 2021;27:e00252. doi:10.1016/j.plabm.2021.e00252
BibTeX
@article{Millius2021,
author = {Millius, Laura and Riedo, Erwin and Caron, Thierry and Belissent, Juliette and Fellay, Benoît and Ribordy, Vincent and Magnin, Jean-Luc},
title = {The „EPiQ“-Study (Evaluation of preanalytical quality): S-Monovette® in manual aspiration mode drastically reduces hemolytic samples in head-to-head study},
journal = {Practical Laboratory Medicine},
year = {2021},
volume = {27},
pages = {e00252},
doi = {10.1016/j.plabm.2021.e00252}
}
EUVIMED