Influence of gating and drift correction on prostate cancer treatment margins on a 1.5 T magnetic resonance linear accelerator
Physics and Imaging in Radiation Oncology · 2026-11 · 42 · 101127
DOI: 10.1016/j.phro.2026.101127 ↗Kurz gefasst
Fachartikel in der Zeitschrift Physics and Imaging in Radiation Oncology, 2026-11. Verfasst von Nienke D. Sijtsema und 4 weiteren Autor:innen. Die Arbeit nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
In this study, intra-fraction motion management using gating and baseline shift corrections on a 1.5 T magnetic resonance linear accelerator was evaluated in 93 prostate cancer patients. Motion tracks were analyzed to assess the frequency, size and duration of baseline shift corrections and derive planning target volume (PTV) margins. Baseline shift corrections were required in 89.2% of patients, with a range of 0–6 corrections per fraction. The resulting PTV margins were 1.2 mm left-right, and 2.3 mm anterior-posterior and cranial-caudal. Because these values were conservatively estimated, a 2 mm isotropic PTV margin was clinically implemented without increasing treatment time.
Zitieren
APA
Sijtsema, N. D., Grimbergen, G., Breugel, E. N. D. G., Zyp, J. R. V. D. V. V., & Boer, J. C. D. (2026). Influence of gating and drift correction on prostate cancer treatment margins on a 1.5 T magnetic resonance linear accelerator. Physics and Imaging in Radiation Oncology, 42, 101127. https://doi.org/10.1016/j.phro.2026.101127
Vancouver
Sijtsema ND, Grimbergen G, Breugel ENDG, Zyp JRVDVV, Boer JCD. Influence of gating and drift correction on prostate cancer treatment margins on a 1.5 T magnetic resonance linear accelerator. Physics and Imaging in Radiation Oncology. 2026;42:101127. doi:10.1016/j.phro.2026.101127
BibTeX
@article{Sijtsema2026,
author = {Sijtsema, Nienke D. and Grimbergen, Guus and Breugel, Eline N. de Groot-van and Zyp, Jochem R.N. van der Voort van and Boer, Johannes C.J. de},
title = {Influence of gating and drift correction on prostate cancer treatment margins on a 1.5 T magnetic resonance linear accelerator},
journal = {Physics and Imaging in Radiation Oncology},
year = {2026},
volume = {42},
pages = {101127},
doi = {10.1016/j.phro.2026.101127}
}
EUVIMED