Implementing a fully computed tomography-free online adaptive palliative radiotherapy: a one-visit workflow

Ashaya T Jaglal, Koen J Nelissen, Angelique R W van Vlaenderen, Amy L de la Fuente, Famke L Schneiders, Peter S N van Rossum, Jan Wiersma, Wilko F A R Verbakel, Suresh Senan, Jorrit Visser, Eva Versteijne

Physics and Imaging in Radiation Oncology · 2026-01 · 37 · 100896

DOI: 10.1016/j.phro.2025.100896 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Physics and Imaging in Radiation Oncology, 2026-01. Verfasst von Ashaya T Jaglal und 10 weiteren Autor:innen. Beteiligte Einrichtung: Amsterdam UMC Location Vrije Universiteit Amsterdam, Department of Radiation Oncology, Amsterdam, the Netherlands. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Varian Medical Systems, Siemens Healthineers.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background and purpose: Same-day palliative radiotherapy requires rapid workflows, but conventional computed tomography (CT)-based workflows cause delays and strain resources. Advances in cone-beam CT (CBCT) enabled accurate dose calculation and planning without a planning CT. This study evaluated the feasibility and efficiency of a fully CT-free online adaptive workflow for same-day palliative radiotherapy using high-quality CBCT. Methods and materials: This prospective study enrolled sixteen patients between January-May 2025, of whom fifteen completed same-day treatment. Eligible patients were referred for single-fraction palliative radiotherapy (8 Gy) to non-mobile target volumes. No planning CT was acquired; instead, a reference plan was generated on a phantom with standardized beam setups and planning objectives. On the treatment day, planning and delivery were performed on the Varian Ethos 2.0 platform using HyperSight CBCT, providing more accurate Hounsfield Unit imaging for automated organs at risk segmentation and target definition. Plans were adapted online and delivered while patients were on the couch. Workflow times, plan quality, and patient characteristics were studied.

Results

All fifteen treatments were delivered successfully. All plans met clinical objectives, with planning target volume coverage exceeding required thresholds. The CT-free workflow reduced median departmental time to 73 min, including 28 min in the treatment room, compared with 335 min in a conventional CT-based workflow. In one urgent case, referral-to-treatment time was 2.5 h. Repeated CBCTs were required in 7 patients.

Conclusions

A fully CT-free workflow for palliative radiotherapy is feasible and efficient, enabling same-day treatment, reduces departmental workload, and is well-suited for urgent cases requiring rapid intervention.

Schlagwörter

Stichwörter der Autor:innen

Cone beam computed tomographyOnline adaptive radiotherapyPalliative care

Autor:innen und Institutionen

  • Amsterdam UMC Location Vrije Universiteit Amsterdam, Department of Radiation Oncology, Amsterdam, the Netherlands.
  • Cancer Center Amsterdam, Cancer Treatment and Quality of Life, Amsterdam, the Netherlands.
  • Varian Medical System, Inc., a Siemens Healthineers Company, Palo Alto, USA.

Förderung

  • Varian Medical Systems
  • Siemens Healthineers

Zitieren

APA

Jaglal, A. T., Nelissen, K. J., van Vlaenderen, A. R. W., de la Fuente, A. L., Schneiders, F. L., van Rossum, P. S. N., Wiersma, J., Verbakel, W. F. A. R., Senan, S., Visser, J., & Versteijne, E. (2026). Implementing a fully computed tomography-free online adaptive palliative radiotherapy: a one-visit workflow. Physics and Imaging in Radiation Oncology, 37, 100896. https://doi.org/10.1016/j.phro.2025.100896

Vancouver

Jaglal AT, Nelissen KJ, van Vlaenderen ARW, de la Fuente AL, Schneiders FL, van Rossum PSN, et al. Implementing a fully computed tomography-free online adaptive palliative radiotherapy: a one-visit workflow. Physics and Imaging in Radiation Oncology. 2026;37:100896. doi:10.1016/j.phro.2025.100896

BibTeX
@article{Jaglal2026,
  author = {Jaglal, Ashaya T and Nelissen, Koen J and van Vlaenderen, Angelique R W and de la Fuente, Amy L and Schneiders, Famke L and van Rossum, Peter S N and Wiersma, Jan and Verbakel, Wilko F A R and Senan, Suresh and Visser, Jorrit and Versteijne, Eva},
  title = {Implementing a fully computed tomography-free online adaptive palliative radiotherapy: a one-visit workflow},
  journal = {Physics and Imaging in Radiation Oncology},
  year = {2026},
  volume = {37},
  pages = {100896},
  doi = {10.1016/j.phro.2025.100896}
}
Kostenlos starten →