Tools and recommendations for commissioning and quality assurance of deformable image registration in radiotherapy
Physics and Imaging in Radiation Oncology · 2024-10 · 32 · 100647
DOI: 10.1016/j.phro.2024.100647 ↗Kurz gefasst
Übersichtsarbeit in der Zeitschrift Physics and Imaging in Radiation Oncology, 2024-10. Verfasst von Lando S Bosma und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiotherapy, University Medical Center Utrecht, Utrecht, the Netherlands. Die Arbeit wurde bisher 14-mal zitiert (Crossref) und nennt 76 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Wellcome / EPSRC Centre for Interventional and Surgical Sciences, Health~Holland, Cancer Research UK.
Abstract (Originalsprache)
Multiple tools are available for commissioning and quality assurance of deformable image registration (DIR), each with their own advantages and disadvantages in the context of radiotherapy. The selection of appropriate tools should depend on the DIR application with its corresponding available input, desired output, and time requirement. Discussions were hosted by the ESTRO Physics Workshop 2021 on Commissioning and Quality Assurance for DIR in Radiotherapy. A consensus was reached on what requirements are needed for commissioning and quality assurance for different applications, and what combination of tools is associated with this. For commissioning, we recommend the target registration error of manually annotated anatomical landmarks or the distance-to-agreement of manually delineated contours to evaluate alignment. These should be supplemented by the distance to discordance and/or biomechanical criteria to evaluate consistency and plausibility. Digital phantoms can be useful to evaluate DIR for dose accumulation but are currently only available for a limited range of anatomies, image modalities and types of deformations. For quality assurance of DIR for contour propagation, we recommend at least a visual inspection of the registered image and contour. For quality assurance of DIR for warping quantitative information such as dose, Hounsfield units or positron emission tomography-data, we recommend visual inspection of the registered image together with image similarity to evaluate alignment, supplemented by an inspection of the Jacobian determinant or bending energy to evaluate plausibility, and by the dose (gradient) to evaluate relevance. We acknowledge that some of these metrics are still missing in currently available commercial solutions.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiotherapy, University Medical Center Utrecht, Utrecht, the Netherlands.
- Metrology for Medical Physics Centre, National Physical Laboratory, Teddington, UK.
- GenesisCare, Sydney, Australia.
- School of Clinical Medicine, Medicine and Health, University of New South Wales, Sydney, Australia.
- Oncology Systems Limited, Shrewsbury, UK.
- Department of Imaging Physics, The University of Texas MD Anderson Cancer Center, Houston, TX, USA.
Förderung
- Wellcome / EPSRC Centre for Interventional and Surgical Sciences
- Health~Holland
- Cancer Research UK
- NWO
Zitieren
APA
Bosma, L. S., Hussein, M., Jameson, M. G., Asghar, S., Brock, K. K., McClelland, J. R., Poeta, S., Yuen, J., Zachiu, C., Yeo, A. U., & 2021 ESTRO Physics Workshop on Commissioning and Quality Assurance for Deformable Image Registration in Radiotherapy (2024). Tools and recommendations for commissioning and quality assurance of deformable image registration in radiotherapy. Physics and Imaging in Radiation Oncology, 32, 100647. https://doi.org/10.1016/j.phro.2024.100647
Vancouver
Bosma LS, Hussein M, Jameson MG, Asghar S, Brock KK, McClelland JR, et al. Tools and recommendations for commissioning and quality assurance of deformable image registration in radiotherapy. Physics and Imaging in Radiation Oncology. 2024;32:100647. doi:10.1016/j.phro.2024.100647
BibTeX
@article{Bosma2024,
author = {Bosma, Lando S and Hussein, Mohammad and Jameson, Michael G and Asghar, Soban and Brock, Kristy K and McClelland, Jamie R and Poeta, Sara and Yuen, Johnson and Zachiu, Cornel and Yeo, Adam U and 2021 ESTRO Physics Workshop on Commissioning and Quality Assurance for Deformable Image Registration in Radiotherapy, 2021 ESTRO Physics Workshop on Commissioning and Quality Assurance for Deformable Image Registration in Radiotherapy},
title = {Tools and recommendations for commissioning and quality assurance of deformable image registration in radiotherapy},
journal = {Physics and Imaging in Radiation Oncology},
year = {2024},
volume = {32},
pages = {100647},
doi = {10.1016/j.phro.2024.100647}
}
EUVIMED