EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Machine learning and lean six sigma for targeted patient-specific quality assurance of volumetric modulated arc therapy plans

Nicola Lambri, Damiano Dei, Giulia Goretti, Leonardo Crespi, Ricardo Coimbra Brioso, Marco Pelizzoli, Sara Parabicoli, Andrea Bresolin, Pasqualina Gallo, Francesco La Fauci, Francesca Lobefalo, Lucia Paganini und 6 weitere

Physics and Imaging in Radiation Oncology · 2024-07 · 31 · 100617

DOI: 10.1016/j.phro.2024.100617 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Physics and Imaging in Radiation Oncology, 2024-07. Verfasst von Nicola Lambri und 17 weiteren Autor:innen. Beteiligte Einrichtung: IRCCS Humanitas Research Hospital, Radiotherapy and Radiosurgery Department, via Manzoni 56, 20089 Rozzano, Milan, Italy. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 43 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Chosun University Dental Hospital, Ministero della Salute, Agenzia Italiana del Farmaco, Ministero della Salute.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background and purpose: Radiotherapy plans with excessive complexity exhibit higher uncertainties and worse patient-specific quality assurance (PSQA) results, while the workload of measurement-based PSQA can impact the efficiency of the radiotherapy workflow. Machine Learning (ML) and Lean Six Sigma, a process optimization method, were implemented to adopt a targeted PSQA approach, aiming to reduce workload, risk of failures, and monitor complexity. Materials and methods: Lean Six Sigma was applied using DMAIC (define, measure, analyze, improve, and control) steps. Ten complexity metrics were computed for 69,811 volumetric modulated arc therapy (VMAT) arcs from 28,612 plans delivered in our Institute (2013-2021). Outlier complexities were defined as >95th-percentile of the historical distributions, stratified by treatment. An ML model was trained to predict the gamma passing rate (GPR-3 %/1mm) of an arc given its complexity. A decision support system was developed to monitor the complexity and expected GPR. Plans at risk of PSQA failure, either extremely complex or with average GPR

Schlagwörter

Stichwörter der Autor:innen

Clinical risk managementLean six sigmaMachine learningQualityRadiotherapy

Autor:innen und Institutionen

  • IRCCS Humanitas Research Hospital, Radiotherapy and Radiosurgery Department, via Manzoni 56, 20089 Rozzano, Milan, Italy.
  • Department of Biomedical Sciences, Humanitas University, via Rita Levi Montalcini 4, 20072 Pieve Emanuele, Milan, Italy.
  • IRCCS Humanitas Research Hospital, Quality Department, via Manzoni 56, 20089 Rozzano, Milan, Italy.
  • Dipartimento di Elettronica, Informazione e Bioingegneria, Politecnico di Milano, 20133 Milan, Italy.
  • Health Data Science Centre, Human Technopole, 20157 Milan, Italy.
  • Dipartimento di Fisica "Aldo Pontremoli", Università degli Studi di Milano, Milan, Italy.

Förderung

  • Chosun University Dental Hospital
  • Ministero della Salute
  • Agenzia Italiana del Farmaco, Ministero della Salute

Zitieren

APA

Lambri, N., Dei, D., Goretti, G., Crespi, L., Brioso, R. C., Pelizzoli, M., Parabicoli, S., Bresolin, A., Gallo, P., La Fauci, F., Lobefalo, F., Paganini, L., Reggiori, G., Loiacono, D., Franzese, C., Tomatis, S., Scorsetti, M., & Mancosu, P. (2024). Machine learning and lean six sigma for targeted patient-specific quality assurance of volumetric modulated arc therapy plans. Physics and Imaging in Radiation Oncology, 31, 100617. https://doi.org/10.1016/j.phro.2024.100617

Vancouver

Lambri N, Dei D, Goretti G, Crespi L, Brioso RC, Pelizzoli M, et al. Machine learning and lean six sigma for targeted patient-specific quality assurance of volumetric modulated arc therapy plans. Physics and Imaging in Radiation Oncology. 2024;31:100617. doi:10.1016/j.phro.2024.100617

BibTeX
@article{Lambri2024,
  author = {Lambri, Nicola and Dei, Damiano and Goretti, Giulia and Crespi, Leonardo and Brioso, Ricardo Coimbra and Pelizzoli, Marco and Parabicoli, Sara and Bresolin, Andrea and Gallo, Pasqualina and La Fauci, Francesco and Lobefalo, Francesca and Paganini, Lucia and Reggiori, Giacomo and Loiacono, Daniele and Franzese, Ciro and Tomatis, Stefano and Scorsetti, Marta and Mancosu, Pietro},
  title = {Machine learning and lean six sigma for targeted patient-specific quality assurance of volumetric modulated arc therapy plans},
  journal = {Physics and Imaging in Radiation Oncology},
  year = {2024},
  volume = {31},
  pages = {100617},
  doi = {10.1016/j.phro.2024.100617}
}
Kostenlos starten →