Parameter based 4D dose calculations for proton therapy
Physics and Imaging in Radiation Oncology · 2023-07 · 27 · 100473
DOI: 10.1016/j.phro.2023.100473 ↗Kurz gefasst
Fachartikel in der Zeitschrift Physics and Imaging in Radiation Oncology, 2023-07. Verfasst von Franciska Lebbink und 6 weiteren Autor:innen. Beteiligte Einrichtung: Medical University of Vienna, Department of Radiation Oncology, Vienna, Austria. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Gesellschaft für Forschungsförderung Niederösterreich.
Abstract (Originalsprache)
Background and purpose Retrospective log file-based analysis provides the actual dose delivered based on the patient's breathing and the daily beam-delivery dynamics. To predict the motion sensitivity of the treatment plan on a patient-specific basis before treatment start a prospective tool is required. Such a parameter-based tool has been investigated with the aim to be used in clinical routine. Materials and Methods 4D dose calculations (4DDC) were performed for seven cancer patients with small breathing motion treated with scanned pulsed proton beams. Validation of the parameter-based 4DDC (p-4DDC) method was performed with an anthropomorphic phantom and patient data employing measurements and a log file-based 4DDC tool. The dose volume histogram parameters (Dx%) were investigated for the target and the organs at risk, compared to static and the file-based approach.
Results
The difference between the measured and the p-4DDC dose was within the deviation of the measurements. The maximum deviation was 0.4Gy. For the planning target volume D98% varied up to 15% compared to the static scenario, while the results from the log file and p-4DDC agreed within 2%. For the liver patients, D33%liver deviated up to 35% compared to static and 10% comparing the two 4DDC tools, while for the pancreas patients the D1%stomach varied up to 45% and 11%, respectively.
Conclusion
The results showed that p-4DDC could be used prospectively. The next step will be the clinical implementation of the p-4DDC tool, which can support a decision to either adapt the treatment plan or apply motion mitigation strategies.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Medical University of Vienna, Department of Radiation Oncology, Vienna, Austria.
- MedAustron Ion Therapy Centre, Wiener Neustadt, Austria.
- Karl Landsteiner University of Health Sciences, Wiener Neustadt, Austria.
- RaySearch Laboratories, Stockholm, Sweden.
Förderung
- Gesellschaft für Forschungsförderung Niederösterreich
Zitieren
APA
Lebbink, F., Stocchiero, S., Fossati, P., Engwall, E., Georg, D., Stock, M., & Knäusl, B. (2023). Parameter based 4D dose calculations for proton therapy. Physics and Imaging in Radiation Oncology, 27, 100473. https://doi.org/10.1016/j.phro.2023.100473
Vancouver
Lebbink F, Stocchiero S, Fossati P, Engwall E, Georg D, Stock M, et al. Parameter based 4D dose calculations for proton therapy. Physics and Imaging in Radiation Oncology. 2023;27:100473. doi:10.1016/j.phro.2023.100473
BibTeX
@article{Lebbink2023,
author = {Lebbink, Franciska and Stocchiero, Silvia and Fossati, Piero and Engwall, Erik and Georg, Dietmar and Stock, Markus and Knäusl, Barbara},
title = {Parameter based 4D dose calculations for proton therapy},
journal = {Physics and Imaging in Radiation Oncology},
year = {2023},
volume = {27},
pages = {100473},
doi = {10.1016/j.phro.2023.100473}
}
EUVIMED