Suction force-suction distance relation during aspiration thrombectomy for ischemic stroke: A computational fluid dynamics study

Yubing Shi, David Cheshire, Frank Lally, Christine Roffe

Physics in Medicine · 2017

DOI: 10.1016/j.phmed.2016.11.001 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Physics in Medicine, 2017. Verfasst von Yubing Shi und 3 weiteren Autor:innen. Die Arbeit wurde bisher 42-mal zitiert (Crossref) und nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Staffordshire University.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Acute Ischemic Stroke (AIS) is the major type of stroke occurring in patients. Aspiration thrombectomy, which uses suction to remove the thrombosis, is a promising technique in the clinical treatment of AIS patients. In this research a computational fluid dynamics (CFD) analysis was conducted to model the blood flow dynamics in a simplified cerebral model during an aspiration thrombectomy procedure. The flow system being analysed was a typical in vitro cerebral flow model, and the system parameters were set based on the clinical and in vitro data reported in open literature. The simulated flow field features showed good correlation with the in vitro response as reported in literature. The CFD study provides detailed technical data including the peak velocity occurring at the catheter tip and the suction force-suction distance relation during the aspiration thrombectomy procedure, which are useful new knowledge and have the potential to influence future catheter design as well as clinical operational protocols used during thrombectomy intervention.

Förderung

  • Staffordshire University

Zitieren

APA

Shi, Y., Cheshire, D., Lally, F., & Roffe, C. (2017). Suction force-suction distance relation during aspiration thrombectomy for ischemic stroke: A computational fluid dynamics study. Physics in Medicine. https://doi.org/10.1016/j.phmed.2016.11.001

Vancouver

Shi Y, Cheshire D, Lally F, Roffe C. Suction force-suction distance relation during aspiration thrombectomy for ischemic stroke: A computational fluid dynamics study. Physics in Medicine. 2017. doi:10.1016/j.phmed.2016.11.001

BibTeX
@article{Shi2017,
  author = {Shi, Yubing and Cheshire, David and Lally, Frank and Roffe, Christine},
  title = {Suction force-suction distance relation during aspiration thrombectomy for ischemic stroke: A computational fluid dynamics study},
  journal = {Physics in Medicine},
  year = {2017},
  doi = {10.1016/j.phmed.2016.11.001}
}
Kostenlos starten →