Fachartikel in der Zeitschrift Annales Pharmaceutiques Françaises, 2026. Verfasst von Clément Beaulieu und 4 weiteren Autor:innen. Beteiligte Einrichtung: Stérilisation, pôle pharmacie, CHU Charles-Nicolle, 1, rue de Germont, 76000 Rouen, France. Electronic address: clmentbeaulieu@gmail.com. Hauptthemen laut MEDLINE: Sterilization. Die Arbeit nennt 12 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Good Hospital Pharmacy Practices require a daily saturated water vapor penetration test to be carried out. In 2021, preliminary work carried out at our facility highlighted the representativeness and sensitivity of Goubanne®. The aim is to roll it out for routine use 24hours a day, 7days a week. METHOD: The project was carried out in three phases. The first phase evaluated the Goubanne® on different sterilization cycles: "rubber", "linen", and "containers". The second phase studied reproducibility on "container" loads and measured the power consumption of BD cycles. Finally, the third phase consisted of training and evaluating staff before the planned implementation in September 2025 and the discontinuation of the BD 3M ETS® test. RESULTS: The Goubanne® is unsuitable for "rubber" and "linen" cycles as the device emerges wet from the autoclaves. On "container" cycles, the compliance rate reached 94.5%. The average power consumption for a BD cycle was 3.17kWh. The questionnaire sent to staff received an average score of 4.93/6. CONCLUSION: Goubanne® appears to be a reliable and more environmentally friendly alternative to the BD 3M ETS® test. It is suitable for "container" cycles, but not for "rubber" and "linen" cycles. Its implementation in September 2025 will require careful monitoring to ensure the safety of the circuit. The assessment of savings, particularly in terms of water, needs to be further explored. This work could serve as a guide for other establishments.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

SterilizationPharmacy Service, HospitalHumansSteamDrug Packaging

Stichwörter der Autor:innen

AutoclavingCentral supply, hospitalEcologyEnvironmental monitoringEnvironnement et santé publiqueEquipment ReuseRéutilisation du matérielService central de stérilisation de l’hôpitalSterilization/standardsStérilisation à l’autoclaveStérilisation/normesÉcologie

Substanzen

Steam

Autor:innen und Institutionen

  • Stérilisation, pôle pharmacie, CHU Charles-Nicolle, 1, rue de Germont, 76000 Rouen, France. Electronic address: clmentbeaulieu@gmail.com.
  • Stérilisation, pôle pharmacie, CHU Charles-Nicolle, 1, rue de Germont, 76000 Rouen, France.

Zitieren

APA

Beaulieu, C., Gombert, C., Machuelle, L., Laurent, M., & Varin, R. (2026). Mise en place du Goubanne® en stérilisation. Annales Pharmaceutiques Françaises. https://doi.org/10.1016/j.pharma.2025.11.006

Vancouver

Beaulieu C, Gombert C, Machuelle L, Laurent M, Varin R. Mise en place du Goubanne® en stérilisation. Annales Pharmaceutiques Françaises. 2026. doi:10.1016/j.pharma.2025.11.006

BibTeX
@article{Beaulieu2026,
  author = {Beaulieu, Clément and Gombert, Cyril and Machuelle, Léa and Laurent, Marc and Varin, Rémi},
  title = {Mise en place du Goubanne® en stérilisation},
  journal = {Annales Pharmaceutiques Françaises},
  year = {2026},
  doi = {10.1016/j.pharma.2025.11.006}
}
Kostenlos starten →