EUVIMEDEUVIMED entdecken
Crossref

Risques associés à l’acquisition d’un automate de dispensation nominative : analyse après dix ans de mise en service au sein d’une pharmacie à usage intérieur

Julie Thiec, Delphine Malet, Anne Colombe, Anne-Laure Debruyne, Emmanuelle Queuille

Annales Pharmaceutiques Françaises · 2025

DOI: 10.1016/j.pharma.2025.05.003 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Annales Pharmaceutiques Françaises, 2025. Verfasst von Julie Thiec und 4 weiteren Autor:innen. Beteiligte Einrichtung: Pharmacie, centre hospitalier Charles-Perrens, 121, rue de la Bechade, CS81285, 33076 Bordeaux cedex, France. Electronic address: thiecjulie@gmail.co…. Hauptthemen laut MEDLINE: Pharmacy Service, Hospital, Robotics, Medication Systems, Hospital. Die Arbeit nennt 23 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

OBJECTIVES: Our hospital pharmaceutical department acquired an automated drug dispensive system to secure the medication management, from the dispensation to the administration process. The aim of our study was to secure the new circuit by identifying the risks associated with the implementation of the robotic system. METHODS: A working-group was set-up and the use of the Failure Modes, Effects and Criticality Analysis (FMECA) method was decided. FMECA is a proactive risk assessment tool that allows the identification of all potential failures. It relies on the identification of failure modes and their severity, frequency and detectability. The product of the three ratings determines the criticality of the failure mode and allows to rank the failure modes according to the criticality: from low, to intermediate and high. RESULTS: The process was delimited from drug analysis to medication delivery to the wards. Six steps and 33 sub-steps were identified. In all, 75 failure modes were identified, 67% of which could lead to a medication error. The working-group identified and prioritised corrective measures to implement. It was decided to focus on the failure modes with a high and intermediate criticality index. CONCLUSIONS: This analysis allowed us to highlight weaknesses is the new process created with the implementation of the robotic system. The sub-steps the more at risk are prescription analysis and pillboxes' manually addings. FMECA is an effective tool easy to implement that allows the pooling of ideas and to be exhaustive in the identification of failure modes.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Pharmacy Service, HospitalRoboticsMedication Systems, HospitalHumansMedication ErrorsRisk Assessment

Stichwörter der Autor:innen

AutomatisationCircuit du médicamentGestion des risquesMedication management processRisk management

Autor:innen und Institutionen

  • Pharmacie, centre hospitalier Charles-Perrens, 121, rue de la Bechade, CS81285, 33076 Bordeaux cedex, France. Electronic address: thiecjulie@gmail.com.
  • Pharmacie, centre hospitalier Charles-Perrens, 121, rue de la Bechade, CS81285, 33076 Bordeaux cedex, France.

Zitieren

APA

Thiec, J., Malet, D., Colombe, A., Debruyne, A. L., & Queuille, E. (2025). Risques associés à l’acquisition d’un automate de dispensation nominative : analyse après dix ans de mise en service au sein d’une pharmacie à usage intérieur. Annales Pharmaceutiques Françaises. https://doi.org/10.1016/j.pharma.2025.05.003

Vancouver

Thiec J, Malet D, Colombe A, Debruyne AL, Queuille E. Risques associés à l’acquisition d’un automate de dispensation nominative : analyse après dix ans de mise en service au sein d’une pharmacie à usage intérieur. Annales Pharmaceutiques Françaises. 2025. doi:10.1016/j.pharma.2025.05.003

BibTeX
@article{Thiec2025,
  author = {Thiec, Julie and Malet, Delphine and Colombe, Anne and Debruyne, Anne-Laure and Queuille, Emmanuelle},
  title = {Risques associés à l’acquisition d’un automate de dispensation nominative : analyse après dix ans de mise en service au sein d’une pharmacie à usage intérieur},
  journal = {Annales Pharmaceutiques Françaises},
  year = {2025},
  doi = {10.1016/j.pharma.2025.05.003}
}
Kostenlos starten →