Multi-mechanism collaboration enhanced photoacoustic analyzer for trace H2S detection
Photoacoustics · 2023-02 · 29 · 100449
DOI: 10.1016/j.pacs.2023.100449 ↗Kurz gefasst
Fachartikel in der Zeitschrift Photoacoustics, 2023-02. Verfasst von Min Guo und 7 weiteren Autor:innen. Beteiligte Einrichtung: School of Optoelectronic Engineering and Instrumentation Science, Dalian University of Technology, Dalian 116024, Liaoning, China. Die Arbeit wurde bisher 30-mal zitiert (Crossref) und nennt 55 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Fundamental Research Funds for the Central Universities, National Natural Science Foundation of China.
Abstract (Originalsprache)
To realize the real-time highly sensitive detection of SF6 decomposition product H2S, a multi-mechanism collaboration enhancement photoacoustic spectroscopy analyzer (MCEPA) based on acoustic resonance enhancement, cantilever enhancement and excitation light enhancement is proposed. An SF6 background gas-induced photoacoustic cell (PAC) was used for acoustic resonance (AR) enhancement of the photoacoustic signals. A fiber-optic acoustic sensor based on a silicon cantilever is optimized and fabricated. The narrow-band acoustic signal enhancement based on cantilever mechanical resonance (MR) is realized in the optimal working frequency band of the PAC. A fiber-coupled DFB cascaded an Erbium-doped fiber amplifier (EDFA) realized the light power enhancement (LPE) of the photoacoustic signals excitation source. Experimental results show that the MR of the fiber-optic silicon cantilever acoustic sensor (FSCAS) is matched with the AR of the PAC and combined with the LPE, which realizes the multi-mechanism collaboration enhancement of weak photoacoustic signals. The Allan-Werle deviation evaluation showed that the minimum detection limit of H2S in the SF6 background is 10.96 ppb when the average time is 200 s. Benefiting from the all-optimization of photoacoustic excitation and detection, the MCEPA has near-field high-sensitivity gas detection capability immune to electromagnetic interference.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- School of Optoelectronic Engineering and Instrumentation Science, Dalian University of Technology, Dalian 116024, Liaoning, China.
Förderung
- Fundamental Research Funds for the Central Universities
- National Natural Science Foundation of China
Zitieren
APA
Guo, M., Zhao, X., Chen, K., Cui, D., Zhang, G., Li, C., Gong, Z., & Yu, Q. (2023). Multi-mechanism collaboration enhanced photoacoustic analyzer for trace H2S detection. Photoacoustics, 29, 100449. https://doi.org/10.1016/j.pacs.2023.100449
Vancouver
Guo M, Zhao X, Chen K, Cui D, Zhang G, Li C, et al. Multi-mechanism collaboration enhanced photoacoustic analyzer for trace H2S detection. Photoacoustics. 2023;29:100449. doi:10.1016/j.pacs.2023.100449
BibTeX
@article{Guo2023,
author = {Guo, Min and Zhao, Xinyu and Chen, Ke and Cui, Dongyu and Zhang, Guangyin and Li, Chenxi and Gong, Zhenfeng and Yu, Qingxu},
title = {Multi-mechanism collaboration enhanced photoacoustic analyzer for trace H2S detection},
journal = {Photoacoustics},
year = {2023},
volume = {29},
pages = {100449},
doi = {10.1016/j.pacs.2023.100449}
}
EUVIMED