Accuracy, Precision, and Residual Volume of Commonly Used Syringes for Intravitreal Injections and the Impact on Intraocular Pressure

Lydianne L M Agra, Alexander Sverstad, Thiago A Chagas, Rodrigo H Araújo, Larissa G Oliveira, Olav Kristianslund, Goran Petrovski, Maurício Maia, Morten Carsten Moe, Øystein Kalsnes Jørstad, Gustavo Barreto Melo

Ophthalmology Retina · 2023

DOI: 10.1016/j.oret.2023.06.003 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Ophthalmology Retina, 2023. Verfasst von Lydianne L M Agra und 10 weiteren Autor:innen. Beteiligte Einrichtung: Hospital de Olhos de Sergipe, Aracaju, Brazil; Department of Ophthalmology, Federal University of São Paulo, São Paulo, Brazil. Hauptthemen laut MEDLINE: Syringes, Eye Diseases. Die Arbeit wurde bisher 21-mal zitiert (Crossref) und nennt 41 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Norges Forskningsråd, Bayer, Allergan.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: To compare accuracy, precision, and residual volume of commonly used syringes for intravitreal injections (IVIs) and to assess the intraocular pressure (IOP) rise by variations in volumes delivered. DESIGN: Experimental laboratory study. SUBJECTS: No subjects were involved in this study. METHODS: We tested 8 syringe models with 2 different needle setups, with 2 different solutions (distilled water or glycerin) and target volumes (50 and 70 μL). To obtain the delivered and residual volumes, we weighed the syringe-needle setups with scale before liquid withdrawal, with liquid, and after liquid release. We also created an experimental eye model to determine the transient rise in IOP following stepwise 10-μL increases in injection volumes. MAIN OUTCOME MEASURES: Delivered and residual volumes, IOP rise. RESULTS: We tested a total of 600 syringe-needle setups. Becton Dickinson (BD) Ultra-Fine (0.34 ± 0.28 μL), Zero Residual (1.53 ± 1.15 μL), and Zero Residual Silicone Oil-free (1.40 ± 1.16 μL) syringes showed the lowest residual volume (P < 0.001) in comparison with the others (range: 24.86 ± 1.78 μL for Injekt-F to 51.97 ± 3.37 μL for Omnifix-F). The most accurate setups were (percentage deviation from target volume): Zero Residual Silicone Oil-free (+ 0.70%), Zero Residual 0.3 ml (+ 4.49%), BD Ultra-Fine (+ 7.83%), Injekt-F (9.42%), Norm-Ject (+ 15.88%), Omnifix-F (+ 16.96%), BD Plastipak Brazil (+17.96%), and BD Plastipak Spain syringes (+ 19.41%). There was a statistically significant difference between the Zero Residual Silicone Oil-free syringe and all other syringes (P < 0.0001), except for the Zero Residual 0.3-ml syringe (P = 0.029). The coefficient of variation was low for all syringes. The modeled IOP rise ranged from 32.3 (standard deviation [SD], 1.4) mmHg for 20-μL injection volume to 76.5 (SD, 1.0) mmHg for 80-μL injection volume. For the standard 50-μL injection volume, the peak pressure was 50.7 (SD, 0.1) mmHg, and the pressure rise duration was 28 (SD, 2) minutes. CONCLUSIONS: There were significant differences in accuracy and residual volume between syringes, whereas they showed a high precision. Volume excess results in a considerable increase in IOP rise after injection. These findings may provide a relevant overview to clinicians and to both device and drug manufacturers regarding pharmacoeconomic, safety, and efficacy issues. FINANCIAL DISCLOSURE(S): Proprietary or commercial disclosure may be found in the Footnotes and Disclosures at the end of this article.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

SyringesEye DiseasesHumansIntravitreal InjectionsResidual VolumeIntraocular PressureSilicone Oils

Stichwörter der Autor:innen

AccuracyAnti-VEGFIntraocular pressureIntravitreal injectionResidual volumeSyringe

Substanzen

Silicone Oils

Autor:innen und Institutionen

  • Hospital de Olhos de Sergipe, Aracaju, Brazil; Department of Ophthalmology, Federal University of São Paulo, São Paulo, Brazil.
  • Department of Ophthalmology, Oslo University Hospital and Institute for Clinical Medicine, Faculty of Medicine, University of Oslo, Oslo, Norway.
  • Instituto de Olhos Cristiano Mendonça, Aracaju, Brazil.
  • Tiradentes University, Aracaju, Brazil.
  • Department of Ophthalmology, Oslo University Hospital and Institute for Clinical Medicine, Faculty of Medicine, University of Oslo, Oslo, Norway; Department of Ophthalmology, University of Split School of Medicine and U…
  • Department of Ophthalmology, Federal University of São Paulo, São Paulo, Brazil.

Förderung

  • Norges Forskningsråd
  • Bayer
  • Allergan

Zitieren

APA

Agra, L. L. M., Sverstad, A., Chagas, T. A., Araújo, R. H., Oliveira, L. G., Kristianslund, O., Petrovski, G., Maia, M., Moe, M. C., Jørstad, Ø. K., & Melo, G. B. (2023). Accuracy, Precision, and Residual Volume of Commonly Used Syringes for Intravitreal Injections and the Impact on Intraocular Pressure. Ophthalmology Retina. https://doi.org/10.1016/j.oret.2023.06.003

Vancouver

Agra LLM, Sverstad A, Chagas TA, Araújo RH, Oliveira LG, Kristianslund O, et al. Accuracy, Precision, and Residual Volume of Commonly Used Syringes for Intravitreal Injections and the Impact on Intraocular Pressure. Ophthalmology Retina. 2023. doi:10.1016/j.oret.2023.06.003

BibTeX
@article{Agra2023,
  author = {Agra, Lydianne L M and Sverstad, Alexander and Chagas, Thiago A and Araújo, Rodrigo H and Oliveira, Larissa G and Kristianslund, Olav and Petrovski, Goran and Maia, Maurício and Moe, Morten Carsten and Jørstad, Øystein Kalsnes and Melo, Gustavo Barreto},
  title = {Accuracy, Precision, and Residual Volume of Commonly Used Syringes for Intravitreal Injections and the Impact on Intraocular Pressure},
  journal = {Ophthalmology Retina},
  year = {2023},
  doi = {10.1016/j.oret.2023.06.003}
}
Kostenlos starten →