Comparison of optical performances of intraocular lenses with different depth of field

Federico Tommasi, Alessio Giorgetti, Caterina Monnetti, Lorenzo Fini, Giovanni Romualdi, Rita Mencucci, Massimo Gurioli, Stefano Cavalieri

Journal of Optometry · 2025

DOI: 10.1016/j.optom.2025.100546 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Journal of Optometry, 2025. Verfasst von Federico Tommasi und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physics and Astronomy, University of Florence, via Giovanni Sansone 1, Fiorentino 50019 Sesto, Italy. Electronic address: federico.tomm…. Hauptthemen laut MEDLINE: Lenses, Intraocular, Optics and Photonics, Depth Perception. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: This study presents a detailed optical characterization of three intraocular lenses (IOLs) comparing two so called "enhanced range of field", Evolux IOL and Tecnis Eyhance IOL and one so called "narrow range of field", a standard monofocal IOL Acrysof IQ. METHODS: The measurements are performed using the PMTF optical bench, basing on the Modulation Transfer Function (MTF) to evaluate the optical performance of each lens. The MTFa, representing the area under the MTF curve, is utilized as a key and synthetic metric to quantify performance across different spatial frequencies, providing insights into the IOLs' behavior as defocus varies. RESULTS: The results highlight the strengths and the weaknesses of the IOLs, with the Evolux showing a broadest depth of focus. The USAF resolution target is used to qualitatively assess the image reproduction at far, intermediate, and near distances, offering a visual representation of the IOLs' capabilities. CONCLUSIONS: These findings provide valuable information for selecting IOLs based on patient-specific visual requirements in cataract and refractive surgery.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Lenses, IntraocularOptics and PhotonicsDepth PerceptionHumansProsthesis DesignVisual AcuityRefraction, Ocular

Stichwörter der Autor:innen

Extended depth of focusIntraocular lensModulation transfer functionOptical bench

Autor:innen und Institutionen

  • Department of Physics and Astronomy, University of Florence, via Giovanni Sansone 1, Fiorentino 50019 Sesto, Italy. Electronic address: federico.tommasi@unifi.it.
  • Department of Physics and Astronomy, University of Florence, via Giovanni Sansone 1, Fiorentino 50019 Sesto, Italy.
  • Eye Clinic, Careggi Hospital, Department of Neurosciences, Psychology, Pharmacology and Child Health (NEUROFARBA), University of Florence, Largo Brambilla 3, 50134 Florence, Italy.

Zitieren

APA

Tommasi, F., Giorgetti, A., Monnetti, C., Fini, L., Romualdi, G., Mencucci, R., Gurioli, M., & Cavalieri, S. (2025). Comparison of optical performances of intraocular lenses with different depth of field. Journal of Optometry. https://doi.org/10.1016/j.optom.2025.100546

Vancouver

Tommasi F, Giorgetti A, Monnetti C, Fini L, Romualdi G, Mencucci R, et al. Comparison of optical performances of intraocular lenses with different depth of field. Journal of Optometry. 2025. doi:10.1016/j.optom.2025.100546

BibTeX
@article{Tommasi2025,
  author = {Tommasi, Federico and Giorgetti, Alessio and Monnetti, Caterina and Fini, Lorenzo and Romualdi, Giovanni and Mencucci, Rita and Gurioli, Massimo and Cavalieri, Stefano},
  title = {Comparison of optical performances of intraocular lenses with different depth of field},
  journal = {Journal of Optometry},
  year = {2025},
  doi = {10.1016/j.optom.2025.100546}
}
Kostenlos starten →