Carotid and aortic plaque imaging using 3D gradient-echo imaging and the three-point Dixon method with improved motion-sensitized driven-equilibrium (iMSDE)

Norikazu Koori, Hiroki Kamekawa, Takehiro Naito, Yasuo Takatsu, Hiraku Fuse, Shin Miyakawa, Kenji Yasue, Masato Takahashi, Kazuma Kurata

Magnetic Resonance Imaging · 2024

DOI: 10.1016/j.mri.2024.03.036 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance Imaging, 2024. Verfasst von Norikazu Koori und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiological Sciences, Ibaraki Prefectural University of Health Sciences 4669-2 Ami, Ibaraki 300-03, Japan; Division of Health Sciences…. Hauptthemen laut MEDLINE: Imaging, Three-Dimensional, Plaque, Atherosclerotic. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 46 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: We devised a method that combines the 3D-Dixon-gradientecho (GRE) method with an improved motion-sensitized driven-equilibrium (iMSDE) to suppress blood flow signals. PURPOSE: The purpose of this study was to evaluate the effectiveness of the new method we developed plaque imaging method (3D-Dixon-GRE with the iMSDE method). STUDY TYPE: Retrospective cohort. POPULATION: Thirty-nine patients who underwent cervical plaque imaging. FIELD STRENGTH/SEQUENCE: 3.0 T/3D-GRE. ASSESSMENT: Signal intensities of the common carotid artery, aorta, plaque, muscle, and subcutaneous fat were measured through the VISTA and the 3D-Dixon-GRE with iMSDE methods, and each contrast was calculated. STATISTICAL TEST: Used the Mann Whitney U test. P-values below 0.05 were considered statistically significant. RESULTS: Plaque and muscle contrast estimated through the VISTA method and 3D-Dixon-GRE with iMSDE method was 1.60 ± 0.96 and 2.04 ± 1.06, respectively, (P < 0.05). The contrast between the flow (common carotid artery and Aorta) and muscle according to the VISTA method and 3D-Dixon-GRE with iMSDE method was 0.24 ± 0.11 and 0.40 ± 0.12, respectively (P < 0.001). Finally, the mean contrast for subcutaneous fat and muscle at six locations was 3.05 ± 1.25 and 0.81 ± 0.23 for the VISTA method and 3D-Dixon-GRE with the iMSDE method, respectively (P < 0.001). DATA CONCLUSION: Compared to the conventional method (VISTA), the 3D-Dixon-GRE with iMSDE method is preferable in relation to the fat suppression effect, but it is disadvantageous regarding blood flow signal suppression. Therefore, the 3D-Dixon-GRE with the iMSDE method could be considered useful for plaque imaging.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Imaging, Three-DimensionalPlaque, AtheroscleroticHumansFemaleMaleAgedRetrospective StudiesMiddle AgedAortaCarotid ArteriesImage Interpretation, Computer-AssistedMagnetic Resonance ImagingMotionAged, 80 and over

Stichwörter der Autor:innen

Aortic plaqueBlack bloodDixonPlaque imagingVISTAiMSDE

Autor:innen und Institutionen

  • Department of Radiological Sciences, Ibaraki Prefectural University of Health Sciences 4669-2 Ami, Ibaraki 300-03, Japan; Division of Health Sciences, Kanazawa University Graduate School of Medical Sciences, 5-11-80 Kod…
  • Department of Radiology, Komaki City Hospital, 1-20 Jyoubushi, Komaki, Aichi 485-8520, Japan.
  • Department of Neurosurgery, Komaki City Hospital, 1-20 Jyoubushi, Komaki, Aichi 485-8520, Japan.
  • Division of Health Sciences, Kanazawa University Graduate School of Medical Sciences, 5-11-80 Kodatsuno, Kanazawa, Ishikawa 920-0942, Japan; School of Medical Sciences, Fujita Health University, Molecular Imaging, 1-98…
  • Department of Radiological Sciences, Ibaraki Prefectural University of Health Sciences 4669-2 Ami, Ibaraki 300-03, Japan.

Zitieren

APA

Koori, N., Kamekawa, H., Naito, T., Takatsu, Y., Fuse, H., Miyakawa, S., Yasue, K., Takahashi, M., & Kurata, K. (2024). Carotid and aortic plaque imaging using 3D gradient-echo imaging and the three-point Dixon method with improved motion-sensitized driven-equilibrium (iMSDE). Magnetic Resonance Imaging. https://doi.org/10.1016/j.mri.2024.03.036

Vancouver

Koori N, Kamekawa H, Naito T, Takatsu Y, Fuse H, Miyakawa S, et al. Carotid and aortic plaque imaging using 3D gradient-echo imaging and the three-point Dixon method with improved motion-sensitized driven-equilibrium (iMSDE). Magnetic Resonance Imaging. 2024. doi:10.1016/j.mri.2024.03.036

BibTeX
@article{Koori2024,
  author = {Koori, Norikazu and Kamekawa, Hiroki and Naito, Takehiro and Takatsu, Yasuo and Fuse, Hiraku and Miyakawa, Shin and Yasue, Kenji and Takahashi, Masato and Kurata, Kazuma},
  title = {Carotid and aortic plaque imaging using 3D gradient-echo imaging and the three-point Dixon method with improved motion-sensitized driven-equilibrium (iMSDE)},
  journal = {Magnetic Resonance Imaging},
  year = {2024},
  doi = {10.1016/j.mri.2024.03.036}
}
Kostenlos starten →