Fallbericht in der Zeitschrift Medical Dosimetry, 2025. Verfasst von Eric Lobb. Beteiligte Einrichtung: Ascension NE Wisconsin St Elizabeth Campus, Appleton, WI 54915, USA. Electronic address: eclobb2@gmail.com. Hauptthemen laut MEDLINE: Brain Neoplasms, Carcinoma, Radiosurgery, Radiotherapy Planning, Computer-Assisted. Die Arbeit nennt 3 Literaturangaben.
Abstract (Originalsprache)
The virtual cone is an innovative MLC-based technique for generating dose distributions comparable to those of physical stereotactic cones. Initially designed for functional radiosurgery applications using a high-definition multileaf collimator (MLC) with 2.5 mm leaf width, this technique has been adapted to a standard 5 mm MLC system for treating small brain metastases. The adapted technique uses preconfigured location-specific control point sequences to produce spherical dose distributions with sharp dose gradients, and facilitates efficient planning through parallelizable template-based workflows. This report highlights the use of the adapted virtual cone technique for treating a 57-year-old patient with a 3 mm brain metastasis from metastatic papillary thyroid carcinoma using a standard multileaf collimator.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Ascension NE Wisconsin St Elizabeth Campus, Appleton, WI 54915, USA. Electronic address: eclobb2@gmail.com.
Zitieren
APA
Lobb, E. (2025). Novel application of an adapted virtual cone technique for the radiosurgical treatment of a 3 mm brain metastasis with a standard multileaf collimator. Medical Dosimetry. https://doi.org/10.1016/j.meddos.2024.10.005
Vancouver
Lobb E. Novel application of an adapted virtual cone technique for the radiosurgical treatment of a 3 mm brain metastasis with a standard multileaf collimator. Medical Dosimetry. 2025. doi:10.1016/j.meddos.2024.10.005
BibTeX
@article{Lobb2025,
author = {Lobb, Eric},
title = {Novel application of an adapted virtual cone technique for the radiosurgical treatment of a 3 mm brain metastasis with a standard multileaf collimator},
journal = {Medical Dosimetry},
year = {2025},
doi = {10.1016/j.meddos.2024.10.005}
}
EUVIMED