EUVIMEDEUVIMED entdecken
ValidierungsstudieOpen Access · CC BY 4.0

Validity and reliability of a novel portable tension-gauge dynamometer for isometric and isotonic seated knee extension strength measurement

Katsumi Takeno, Victor L Katch, Grant E Norte, Christopher D Ingersoll

The Knee · 2025

DOI: 10.1016/j.knee.2025.07.013 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift The Knee, 2025. Verfasst von Katsumi Takeno und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Kinesiology, University of North Georgia, Dahlonega, GA, USA. Electronic address: Katsumi.takeno@ung.edu. Hauptthemen laut MEDLINE: Muscle Strength Dynamometer, Isometric Contraction, Muscle Strength, Knee Joint. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: University of North Georgia.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Accessible measurements of multimodal strength are essential to clinical practice and outcome research. While isokinetic dynamometers are considered the gold standard for isometric and isokinetic strength measurement, their practical use is limited by the insufficiency to measure isotonic strength, high cost, and needs for extensive physical space. Establishing techniques to assess isotonic strength using a portable dynamometer may enhance performance evaluations. Our objective was to determine the validity and test-retest reliability of isometric and isotonic knee extension strength measurements using a novel portable tension-gauge dynamometer. METHODS: Twenty recreationally active college-aged individuals participated. We recorded knee extension maximal voluntary isometric contraction (MVIC) force using both a tension-gauge dynamometer and an isokinetic dynamometer for validity and reliability. Peak isotonic knee extension force was recorded against external loads equivalent to 25 % and 30 % MVIC using the tension-gauge dynamometer for test-retest reliability during two sessions 7 days apart. RESULTS: We observed a very high correlation in the MVICs obtained between devices (r = 0.98, p < 0.001, R2 = 0.96).We observed good-to-excellent reliability for MVIC force (ICC3,1 = 0.94 [0.86, 0.98], p = 0.01), good-to-excellent reliability for peak isotonic force against external load at 25 % MVIC (ICC3,1 = 0.95 [0.87, 0.98], p < 0.001), and excellent reliability for peak isotonic force against external load at 30 % MVIC (ICC3,1 = 0.97 [0.93, 0.99], p < 0.001). CONCLUSIONS: Valid and reliable force-based outcomes can be measured using a novel, tension-gauge dynamometer, which may provide relevant characteristics of muscle function for clinical and research uses.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Muscle Strength DynamometerIsometric ContractionMuscle StrengthKnee JointIsotonic ContractionHumansReproducibility of ResultsMaleYoung AdultFemaleAdultEquipment Design

Stichwörter der Autor:innen

Contraction velocityIsotonic contractionMaximum voluntary isometric contractionMuscle functionStrength measures

Autor:innen und Institutionen

  • Department of Kinesiology, University of North Georgia, Dahlonega, GA, USA. Electronic address: Katsumi.takeno@ung.edu.
  • Department of Kinesiology, University of Michigan, Ann Arbor, MI, USA.
  • Cognition, Neuroplasticity & Sarcopenia (CNS) Laboratory, Institute of Exercise Physiology and Rehabilitation Science, University of Central Florida, Orlando, FL, USA.
  • School of Medicine, University of North Carolina Chapel Hill, Chapel Hill, NC, USA.

Förderung

  • University of North Georgia

Zitieren

APA

Takeno, K., Katch, V. L., Norte, G. E., & Ingersoll, C. D. (2025). Validity and reliability of a novel portable tension-gauge dynamometer for isometric and isotonic seated knee extension strength measurement. The Knee. https://doi.org/10.1016/j.knee.2025.07.013

Vancouver

Takeno K, Katch VL, Norte GE, Ingersoll CD. Validity and reliability of a novel portable tension-gauge dynamometer for isometric and isotonic seated knee extension strength measurement. The Knee. 2025. doi:10.1016/j.knee.2025.07.013

BibTeX
@article{Takeno2025,
  author = {Takeno, Katsumi and Katch, Victor L and Norte, Grant E and Ingersoll, Christopher D},
  title = {Validity and reliability of a novel portable tension-gauge dynamometer for isometric and isotonic seated knee extension strength measurement},
  journal = {The Knee},