EUVIMEDEUVIMED entdecken
Fallbericht

A rare case of taper junction corrosion in semi-constrained total knee arthroplasty

Nnadozie Ekweariri, Ryan White, Nicholas Brown, Daniel Schmitt

The Knee · 2024

DOI: 10.1016/j.knee.2024.02.010 ↗

Kurz gefasst

Fallbericht in der Zeitschrift The Knee, 2024. Verfasst von Nnadozie Ekweariri und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedic Surgery, Loyola University Medical Center, Maywood, IL, USA. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Knee, Knee Prosthesis, Prosthesis Failure, Reoperation. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 24 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Metallosis is a known yet rare late complication of unicompartmental and total knee arthroplasty (TKA), usually secondary to either metal-backed patellar component failure, mobile-bearing polyethylene dislocation, or catastrophic polyethylene failure and wear through. The majority of literature surrounding metallosis has been published in relation to total hip arthroplasty (THA) metal on metal bearing wear or mechanically assisted crevice corrosion.This case report describes the development of metallosis in a 77-year-old male patient with advanced (Kellgren-Lawrence Grade 4) osteoarthritis with associated valgus deformity, who underwent index TKA with a semiconstrained revision knee system due to intraoperative medial collateral ligament laxity. The taper junction between the titanium alloy stem and cobalt chromium femoral component was the source of diffuse intra-articular metallosis.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Arthroplasty, Replacement, KneeKnee ProsthesisProsthesis FailureReoperationHumansMaleAgedCorrosionOsteoarthritis, KneeProsthesis DesignTitanium

Stichwörter der Autor:innen

MetallosisRevisionTotal knee arthroplastyTrunnionosis

Substanzen

Titanium

Autor:innen und Institutionen

  • Department of Orthopaedic Surgery, Loyola University Medical Center, Maywood, IL, USA.

Zitieren

APA

Ekweariri, N., White, R., Brown, N., & Schmitt, D. (2024). A rare case of taper junction corrosion in semi-constrained total knee arthroplasty. The Knee. https://doi.org/10.1016/j.knee.2024.02.010

Vancouver

Ekweariri N, White R, Brown N, Schmitt D. A rare case of taper junction corrosion in semi-constrained total knee arthroplasty. The Knee. 2024. doi:10.1016/j.knee.2024.02.010

BibTeX
@article{Ekweariri2024,
  author = {Ekweariri, Nnadozie and White, Ryan and Brown, Nicholas and Schmitt, Daniel},
  title = {A rare case of taper junction corrosion in semi-constrained total knee arthroplasty},
  journal = {The Knee},
  year = {2024},
  doi = {10.1016/j.knee.2024.02.010}
}
Kostenlos starten →