Low-profile endografts are associated with increased risk of type I/III endoleaks at midterm follow-up of 1220 physician-modified fenestrated, branched endovascular repairs for complex abdominal and thoracoabdominal aortic aneurysms
Journal of Vascular Surgery · 2026
DOI: 10.1016/j.jvs.2025.08.002 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift Journal of Vascular Surgery, 2026. Verfasst von Sukgu M Han und 12 weiteren Autor:innen. Beteiligte Einrichtung: Keck Medical Center of University of Southern California, Los Angeles, CA. Electronic address: sukgu.han@med.usc.edu. Hauptthemen laut MEDLINE: Aortic Aneurysm, Thoracic, Endovascular Procedures, Blood Vessel Prosthesis, Endoleak. Die Arbeit nennt 18 Literaturangaben. Gefördert von: University of California, San Diego, University of Florida, University of Texas Southwestern Medical Center.
Abstract (Originalsprache)
OBJECTIVE: Physician-modified endografts (PMEGs) are widely used for complex abdominal and thoracoabdominal aortic repairs, offering versatility in patient-specific stent graft configurations. Low-profile (LP) endografts improve deliverability and decrease access site complications compared with standard-profile (SP) devices. However, concerns have been raised about the long-term durability of certain LP devices. Although LP devices have been incorporated into PMEGs at many centers, data on their impact on outcomes are limited. Therefore, this study compared outcomes of LP and SP endografts used with PMEGs using an international multicenter database. METHODS: A retrospective analysis of data on patients who received PMEGs at 19 international centers from 2009 to 2022 was conducted. Patients were grouped by LP or SP aortic endografts used for modification. Baseline patient characteristics, aortic pathologies, modification techniques, and outcomes were compared. Primary outcomes included technical success, 30-day mortality, and type I and III endoleaks. The secondary outcomes were major adverse events, access site complications, all-cause and aneurysm-related mortality, and target vessel instability during follow-up. RESULTS: Among 1220 patients included (393 LP, 827 SP), the LP group was older and had higher rates of congestive heart failure, any history of smoking, and an American Society of Anesthesiologists risk score of ≥3. LP devices were more commonly used in elective cases; with short neck aneurysms, postdissection aneurysms, and prior endovascular repairs; and during the latter one-half of the study period. Reinforced fenestration was the most common modification in both groups, with directional branches and preloaded components more common in the LP groups. The LP group required fewer access conduits and a lower radiation dose. There was no difference in technical success (92.5 LP vs 94.5% SP; P = .19), mortality (4.1 LP vs 6.8% LP; P = .06), major adverse events (23.1 LP vs 25.7% SP; P = .38), and access-site complications (8.1 LP vs 11.7% SP; P = .06) at 30 days as well as all-cause and aneurysm-related mortality and target vessel instability during follow-up. However, at a median follow-up of 20 months, significantly more type I or III endoleaks (23% LP vs 13% SP; P < .001) were seen in the LP group. Furthermore, the estimated 2-year branch-related endoleaks (type Ic, IIIc) were higher (hazard ratio, 1.65; P = .014) in the LP group. CONCLUSIONS: The use of LP devices facilitated safe PMEG repair with a lesser need for access conduit. However, the higher rates of type I, type III, and branch-related endoleaks with LP devices raise concerns about their long-term durability; however, more study is required. Based on these data, the use of LP devices for PMEGs should be approached with caution.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Keck Medical Center of University of Southern California, Los Angeles, CA. Electronic address: sukgu.han@med.usc.edu.
- Centro Hospitalar Universitário Lisboa Norte, Lisbon, Portugal.
- University of Alabama, Birmingham, AL.
- Beth Israel Deaconess Medical Center, Boston, MA.
- German Aortic Center Hamburg, Hamburg, Germany.
- University of Washington, Seattle, WA.
Förderung
- University of California, San Diego
- University of Florida
- University of Texas Southwestern Medical Center
- University Hospitals Birmingham NHS Foundation Trust
- University of Texas at Austin
- University of Washington
Zitieren
APA
Han, S. M., Melo, R. G., Beck, A., Schermerhorn, M., Kölbel, T., Sweet, M., Barleban, A., Adam, D., Farber, M., Mendes, B., Oderich, G., Tsilimparis, N., & International Multicenter PMEG Study Group (2026). Low-profile endografts are associated with increased risk of type I/III endoleaks at midterm follow-up of 1220 physician-modified fenestrated, branched endovascular repairs for complex abdominal and thoracoabdominal aortic aneurysms. Journal of Vascular Surgery. https://doi.org/10.1016/j.jvs.2025.08.002
Vancouver
Han SM, Melo RG, Beck A, Schermerhorn M, Kölbel T, Sweet M, et al. Low-profile endografts are associated with increased risk of type I/III endoleaks at midterm follow-up of 1220 physician-modified fenestrated, branched endovascular repairs for complex abdominal and thoracoabdominal aortic aneurysms. Journal of Vascular Surgery. 2026. doi:10.1016/j.jvs.2025.08.002
BibTeX
@article{Han2026,
author = {Han, Sukgu M and Melo, Ryan Gouveia and Beck, Adam and Schermerhorn, Marc and Kölbel, Tilo and Sweet, Matthew and Barleban, Andrew and Adam, Donald and Farber, Mark and Mendes, Bernardo and Oderich, Gustavo and Tsilimparis, Nikolaos and International Multicenter PMEG Study Group, International Multicenter PMEG Study Group},
title = {Low-profile endografts are associated with increased risk of type I/III endoleaks at midterm follow-up of 1220 physician-modified fenestrated, branched endovascular repairs for complex abdominal and thoracoabdominal aortic aneurysms},
journal = {Journal of Vascular Surgery},
year = {2026},
doi = {10.1016/j.jvs.2025.08.002}
}
EUVIMED