Percutaneous CT-Guided Abdominal and Pelvic Biopsies: Comparison of an Electromagnetic Navigation System and CT Fluoroscopy

Annie M Zlevor, Meridith A Kisting, Allison B Couillard, Annika E Rossebo, Timothy P Szczykutowicz, Lu Mao, James K White, Michael P Hartung, Lori Mankowski Gettle, J Louis Hinshaw, Perry J Pickhardt, Timothy J Ziemlewicz und 2 weitere

Journal of Vascular and Interventional Radiology · 2023-05 · 34 (5) · 910-918

DOI: 10.1016/j.jvir.2023.01.024 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Vascular and Interventional Radiology, 2023-05. Verfasst von Annie M Zlevor und 13 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, University of Wisconsin-Madison, Madison, Wisconsin. Hauptthemen laut MEDLINE: Electromagnetic Phenomena, Tomography, X-Ray Computed. Die Arbeit wurde bisher 19-mal zitiert (Crossref) und nennt 36 Literaturangaben. Gefördert von: NCATS NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: To compare electromagnetic navigation (EMN) with computed tomography (CT) fluoroscopy for guiding percutaneous biopsies in the abdomen and pelvis. MATERIALS AND METHODS: A retrospective matched-cohort design was used to compare biopsies in the abdomen and pelvis performed with EMN (consecutive cases, n = 50; CT-Navigation; Imactis, Saint-Martin-d'Hères, France) with those performed with CT fluoroscopy (n = 100). Cases were matched 1:2 (EMN:CT fluoroscopy) for target organ and lesion size (±10 mm). RESULTS: The population was well-matched (age, 65 vs 65 years; target size, 2.0 vs 2.1 cm; skin-to-target distance, 11.4 vs 10.7 cm; P > .05, EMN vs CT fluoroscopy, respectively). Technical success (98% vs 100%), diagnostic yield (98% vs 95%), adverse events (2% vs 5%), and procedure time (33 minutes vs 31 minutes) were not statistically different (P > .05). Operator radiation dose was less with EMN than with CT fluoroscopy (0.04 vs 1.2 μGy; P < .001), but patient dose was greater (30.1 vs 9.6 mSv; P < .001) owing to more helical scans during EMN guidance (3.9 vs 2.1; P < .001). CT fluoroscopy was performed with a mean of 29.7 tap scans per case. In 3 (3%) cases, CT fluoroscopy was performed with gantry tilt, and the mean angle out of plane for EMN cases was 13.4°. CONCLUSIONS: Percutaneous biopsies guided by EMN and CT fluoroscopy were closely matched for technical success, diagnostic yield, procedure time, and adverse events in a matched cohort of patients. EMN cases were more likely to be performed outside of the gantry plane. Radiation dose to the operator was higher with CT fluoroscopy, and patient radiation dose was higher with EMN. Further study with a wider array of procedures and anatomic locations is warranted.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Electromagnetic PhenomenaTomography, X-Ray ComputedHumansAgedRetrospective StudiesBiopsyAbdomenPelvisFluoroscopy

Autor:innen und Institutionen

  • Department of Radiology, University of Wisconsin-Madison, Madison, Wisconsin.
  • Department of Surgery, Medical College of Wisconsin, Milwaukee, Wisconsin.
  • Department of Radiology, University of Wisconsin-Madison, Madison, Wisconsin; Department of Medical Physics, University of Wisconsin-Madison, Madison, Wisconsin; Department of Biomedical Engineering, University of Wisco…
  • Department of Biostatistics and Medical Informatics, University of Wisconsin-Madison, Madison, Wisconsin.
  • Department of Radiology, University of Wisconsin-Madison, Madison, Wisconsin; Department of Urology, University of Wisconsin-Madison, Madison, Wisconsin.
  • Department of Radiology, University of Wisconsin-Madison, Madison, Wisconsin; Department of Biomedical Engineering, University of Wisconsin-Madison, Madison, Wisconsin; Department of Urology, University of Wisconsin-Mad…

Förderung

  • NCATS NIH HHS

Zitieren

APA

Zlevor, A. M., Kisting, M. A., Couillard, A. B., Rossebo, A. E., Szczykutowicz, T. P., Mao, L., White, J. K., Hartung, M. P., Gettle, L. M., Hinshaw, J. L., Pickhardt, P. J., Ziemlewicz, T. J., Foltz, M. L., & Lee, F. T. (2023). Percutaneous CT-Guided Abdominal and Pelvic Biopsies: Comparison of an Electromagnetic Navigation System and CT Fluoroscopy. Journal of Vascular and Interventional Radiology, 34(5), 910-918. https://doi.org/10.1016/j.jvir.2023.01.024

Vancouver

Zlevor AM, Kisting MA, Couillard AB, Rossebo AE, Szczykutowicz TP, Mao L, et al. Percutaneous CT-Guided Abdominal and Pelvic Biopsies: Comparison of an Electromagnetic Navigation System and CT Fluoroscopy. Journal of Vascular and Interventional Radiology. 2023;34(5):910-918. doi:10.1016/j.jvir.2023.01.024

BibTeX
@article{Zlevor2023,
  author = {Zlevor, Annie M and Kisting, Meridith A and Couillard, Allison B and Rossebo, Annika E and Szczykutowicz, Timothy P and Mao, Lu and White, James K and Hartung, Michael P and Gettle, Lori Mankowski and Hinshaw, J Louis and Pickhardt, Perry J and Ziemlewicz, Timothy J and Foltz, Marcia L and Lee, Fred T},
  title = {Percutaneous CT-Guided Abdominal and Pelvic Biopsies: Comparison of an Electromagnetic Navigation System and CT Fluoroscopy},
  journal = {Journal of Vascular and Interventional Radiology},
  year = {2023},
  volume = {34},
  number = {5},
  pages = {910--918},
  doi = {10.1016/j.jvir.2023.01.024}
}
Kostenlos starten →