EUVIMEDEUVIMED entdecken
Fallbericht

Insulation break of the ventricular lead in a dog with a pacemaker

H W Jeong, G Menciotti

Journal of Veterinary Cardiology · 2023

DOI: 10.1016/j.jvc.2023.09.005 ↗

Kurz gefasst

Fallbericht in der Zeitschrift Journal of Veterinary Cardiology, 2023. Verfasst von H W Jeong und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department Small Animal Clinical Science, Virginia Maryland College of Veterinary Medicine, Blacksburg, VA, USA. Electronic address: hjeong@northstar…. Hauptthemen laut MEDLINE: Pacemaker, Artificial, Dog Diseases. Die Arbeit nennt 17 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

A 10-year-old female spayed mixed breed dog that had a transvenous pacemaker implanted four years prior was presented for the evaluation of episodes of transient loss of consciousness without any history of trauma that were unresponsive to anticonvulsant therapies. Intermittent failure to capture coupled with low impedance and high capture threshold were noted upon interrogation of the pacemaker. Radiography and fluoroscopy demonstrated a possible break in the insulation of the lead. After the placement of another transvenous pacemaker, the dog had no additional syncopal episodes, and anticonvulsants were tapered off. The dog developed an implant infection from the newly placed pacemaker, which prompted the removal of the infected pacemaker and the implantation of an epicardial pacemaker. However, the epicardial lead dislodged seven days after placement, and the dog was euthanized. During necropsy, the first transvenous lead was removed, and the insulation break was visually confirmed. Although an insulation break is a rare delayed complication of pacemaker implantation, this report shows how the diagnosis, suspected by typical interrogation findings, can be aided by radiographic and fluoroscopic investigations.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Pacemaker, ArtificialDog DiseasesFemaleDogsAnimalsPostoperative Complications

Stichwörter der Autor:innen

Atrioventricular blockBradyarrhythmiaCanine

Autor:innen und Institutionen

  • Department Small Animal Clinical Science, Virginia Maryland College of Veterinary Medicine, Blacksburg, VA, USA. Electronic address: hjeong@northstarvets.com.
  • Department Small Animal Clinical Science, Virginia Maryland College of Veterinary Medicine, Blacksburg, VA, USA.

Zitieren

APA

Jeong, H. W., & Menciotti, G. (2023). Insulation break of the ventricular lead in a dog with a pacemaker. Journal of Veterinary Cardiology. https://doi.org/10.1016/j.jvc.2023.09.005

Vancouver

Jeong HW, Menciotti G. Insulation break of the ventricular lead in a dog with a pacemaker. Journal of Veterinary Cardiology. 2023. doi:10.1016/j.jvc.2023.09.005