EUVIMEDEUVIMED entdecken
Multizentrische StudieOpen Access · CC BY-NC-ND 4.0

Clinical and radiographic survival rates of the Arthrex Univers VaultLock uncemented central-peg all-polyethylene glenoid component in anatomic total shoulder arthroplasty at 2- and 5-years postoperatively

Ehab M Nazzal, Ryan T Lin, Sophia McMahon, Shoulder Arthroplasty Research Committee (ShARC), Patrick J Denard, Kevin Farmer, G Russell Huffman, Albert Lin

Journal of Shoulder and Elbow Surgery · 2026-12 · 35 (12) · e1095-e1107

DOI: 10.1016/j.jse.2026.05.021 ↗

Kurz gefasst

Multizentrische Studie in der Zeitschrift Journal of Shoulder and Elbow Surgery, 2026-12. Verfasst von Ehab M Nazzal und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedic Surgery, University of Pittsburgh Medical Center, Pittsburgh, PA, USA. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Shoulder, Shoulder Prosthesis, Shoulder Joint. Die Arbeit nennt 58 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Arthrex.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: To evaluate radiographic and functional outcomes at 2 and 5 years postoperatively for anatomic total shoulder arthroplasty (aTSA) performed with an uncemented central-peg all-polyethylene glenoid (Univers VaultLock; Arthrex, Inc., Naples, FL) designed for an uncemented central-peg fixation with peripheral cement and central osseous interdigitation. METHODS: This was a multicenter retrospective case series of aTSAs identified from a prospectively maintained registry. All primary aTSAs with an intact rotator cuff using a VaultLock glenoid (uncemented central-peg all-polyethylene design) and available radiographic data were included for analysis at 2- and 5-years postoperatively. Primary outcomes included radiographic glenoid loosening (via Lazarus classification) and revision rates for glenoid loosening. Relevant changes in lucency were defined as change in Lazarus score ≥2 and an "at-risk" component for loosening was defined as Lazarus score ≥3. Secondary outcomes included range of motion, as well as patient-reported outcomes. RESULTS: A total of 133 patients were included in the overall cohort, of whom 119 met criteria for analysis of radiographic and patient-reported outcomes at 5 years. The mean age was 63.9 ± 7.8 years, with an average body mass index of 30.2 ± 4.9 kg/m2 and mean follow-up of 5.0 ± 0.4 years. All outcome measures significantly improved at 2- and 5-years compared to baseline. At 2 years, 111 patients (93.3%) demonstrated a grade 0-1 Lazarus score, while 8 patients (6.7%) had a grade 2 Lazarus score. No glenoids at 2 years were observed to have grade 3 changes consistent with being "at-risk" for radiographic loosening. At 5 years, 100 patients (84.0%) had a grade 0-1 Lazarus score, 13 patients (11.0%) had a grade 2 Lazarus score, and 6 patients (5.0%) had a grade 3 or greater Lazarus scores. Forty (33.6%) patients had an increase in Lazarus score from 2 to 5 years with mean increase of 1.2 (SD 0.42, range: 1-2). At 5 years, central peg humeral head bone grafting was associated with lower rates of Lazarus score ≥2 vs. no grafting (9.5% vs. 23.2%, P = .042). Clinical survival rate was 99.2% (132/133) at 5-years postoperatively, with radiographic survival of 95.0% (113/119) at 5-years postoperatively. CONCLUSIONS: The VaultLock glenoid has a 5-year clinical and radiographic glenoid survival rate of 99.2% and 95.0%, respectively. These findings are consistent with or superior to prior published reports of all-polyethylene glenoids and support the use of an uncemented central-peg all-polyethylene component in primary TSA.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Arthroplasty, Replacement, ShoulderShoulder ProsthesisShoulder JointHumansRetrospective StudiesProsthesis DesignFemaleProsthesis FailureMaleRange of Motion, ArticularAgedRadiographyMiddle AgedPolyethyleneFollow-Up StudiesTime FactorsTreatment Outcome

Stichwörter der Autor:innen

Anatomic total shoulder arthroplastyFunctional outcomesGlenoid looseningRadiographic outcomesUncemented central-peg all-polyethylene glenoidVaultLock glenoid

Substanzen

Polyethylene

Autor:innen und Institutionen

  • Department of Orthopaedic Surgery, University of Pittsburgh Medical Center, Pittsburgh, PA, USA.
  • Oregon Shoulder Institute, Medford, OR, USA.
  • Department of Orthopedic Surgery and Sports Medicine, University of Florida, Gainesville, FL, USA.
  • Rothman Orthopaedics Florida at AdventHealth, Orlando, FL, USA.
  • Department of Orthopaedic Surgery, University of Pittsburgh Medical Center, Pittsburgh, PA, USA. Electronic address: lina2@upmc.edu.

Förderung

  • Arthrex

Zitieren

APA

Nazzal, E. M., Lin, R. T., McMahon, S., Shoulder Arthroplasty Research Committee (ShARC), Denard, P. J., Farmer, K., Huffman, G. R., & Lin, A. (2026). Clinical and radiographic survival rates of the Arthrex Univers VaultLock uncemented central-peg all-polyethylene glenoid component in anatomic total shoulder arthroplasty at 2- and 5-years postoperatively. Journal of Shoulder and Elbow Surgery, 35(12), e1095-e1107. https://doi.org/10.1016/j.jse.2026.05.021

Vancouver

Nazzal EM, Lin RT, McMahon S, Shoulder Arthroplasty Research Committee (ShARC), Denard PJ, Farmer K, et al. Clinical and radiographic survival rates of the Arthrex Univers VaultLock uncemented central-peg all-polyethylene glenoid component in anatomic total shoulder arthroplasty at 2- and 5-years postoperatively. Journal of Shoulder and Elbow Surgery. 2026;35(12):e1095-e1107. doi:10.1016/j.jse.2026.05.021

BibTeX
@article{Nazzal2026,
  author = {Nazzal, Ehab M and Lin, Ryan T and McMahon, Sophia and Shoulder Arthroplasty Research Committee (ShARC), Shoulder Arthroplasty Research Committee (ShARC) and Denard, Patrick J and Farmer, Kevin and Huffman, G Russell and Lin, Albert},
  title = {Clinical and radiographic survival rates of the Arthrex Univers VaultLock uncemented central-peg all-polyethylene glenoid component in anatomic total shoulder arthroplasty at 2- and 5-years postoperatively},
  journal = {Journal of Shoulder and Elbow Surgery},
  year = {2026},
  volume = {35},
  number = {12},
  pages = {e1095--e1107},
  doi = {10.1016/j.jse.2026.05.021}
}
Kostenlos starten →