Unicortical button fixation provides higher strength compared with transosseous repair for subscapularis tendon in total shoulder arthroplasty
Journal of Shoulder and Elbow Surgery · 2025
DOI: 10.1016/j.jse.2024.07.019 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Journal of Shoulder and Elbow Surgery, 2025. Verfasst von Phob Ganokroj und 9 weiteren Autor:innen. Beteiligte Einrichtung: The Steadman Philippon Research Institute, Vail, CO, USA; Faculty of Medicine Siriraj Hospital, Mahidol University, Bangkok, Thailand. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Shoulder, Shoulder Joint, Tendon Injuries, Rotator Cuff. Die Arbeit nennt 35 Literaturangaben. Gefördert von: Arthrex.
Abstract (Originalsprache)
BACKGROUND: Subscapularis tendon (SSc) dysfunction after total shoulder arthroplasty (TSA) results in poor functional outcomes. There have been numerous SSc repair constructs tested biomechanically and clinically; however, none has been demonstrated as superior. Newer techniques and implants have emerged but have not been fully tested. HYPOTHESIS: We hypothesized that the unicortical button (UB) fixation would provide significantly improved restoration of the anatomic footprint and biomechanical properties compared with transosseous (TO) repair of the SSc. METHODS: A digital footprint of SSc humeral insertion was obtained in 6 pairs of fresh-frozen cadaveric shoulders using a 3-dimensional digitizer. A complete SSc tear was created, and each pair of shoulders was randomized to either SSc repair with UB or TO repair. Each specimen underwent a cyclic loading protocol, followed by pull to failure. The failure load, elongation at failure, gapping failure, number of cycles until failure, the load at key gapping points (1 mm, 3 mm, 5 mm, and 10 mm), and the failure mode were recorded using high-resolution video recording. Three-dimensional surfaces of the insertion footprint and repair site were obtained, and surface areas were calculated using a custom MATLAB script and laser scanner. Paired t tests were conducted to compare differences between the 2 repair groups. RESULTS: Failure load was significantly higher in the UB group (382.4 N ± 56.5 N) than in the TO group (253.6 N ± 103.4 N, P = .005). TO repair provided higher gapping at failure (28.8 mm ± 8.2 mm) than UB repair (10.4 mm ± 6.8 mm, P = .0017). UB repair had significantly higher load at the 1-mm, 5-mm, and 10-mm gapping than TO repair, with P = .042, P = .033, and P = .0076, respectively. There were no significant differences in elongation failure, the difference in footprint area from native to repair states, or the percentage of the restored footprint area between the groups (P = .26, P = .18, and P = .21, respectively). CONCLUSIONS: The UB fixation showed a significantly lower gap at failure, higher failure load and number of cycles until failure, and higher gap loads than the traditional TO repair for SSc. Although more clinical research is necessary, the UB fixation that uses cortical bone presents promising results.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- The Steadman Philippon Research Institute, Vail, CO, USA; Faculty of Medicine Siriraj Hospital, Mahidol University, Bangkok, Thailand.
- The Steadman Philippon Research Institute, Vail, CO, USA.
- The Steadman Philippon Research Institute, Vail, CO, USA; Department of Orthopaedic Surgery, Brooke Army Medical Center, Fort Sam Houston, TX, USA.
- Institute of Orthopaedics, Lerdsin General Hospital, Bangkok, Thailand.
- The Steadman Philippon Research Institute, Vail, CO, USA; Department of Orthopaedic Surgery, Michigan State University, East Lansing, MI, USA.
- The Steadman Philippon Research Institute, Vail, CO, USA; The Steadman Clinic, Vail, CO, USA. Electronic address: mprovencher@thesteadmanclinic.com.
Förderung
- Arthrex
Zitieren
APA
Ganokroj, P., Garcia, A. R., Hollenbeck, J. F. M., Whalen, R. J., Brown, J. R., Drumm, A., McBride, T. J., Suppauksorn, S., Jildeh, T. R., & Provencher, M. T. (2025). Unicortical button fixation provides higher strength compared with transosseous repair for subscapularis tendon in total shoulder arthroplasty. Journal of Shoulder and Elbow Surgery. https://doi.org/10.1016/j.jse.2024.07.019
Vancouver
Ganokroj P, Garcia AR, Hollenbeck JFM, Whalen RJ, Brown JR, Drumm A, et al. Unicortical button fixation provides higher strength compared with transosseous repair for subscapularis tendon in total shoulder arthroplasty. Journal of Shoulder and Elbow Surgery. 2025. doi:10.1016/j.jse.2024.07.019
BibTeX
@article{Ganokroj2025,
author = {Ganokroj, Phob and Garcia, Alexander R and Hollenbeck, Justin F M and Whalen, Ryan J and Brown, Justin R and Drumm, Amelia and McBride, Trevor J and Suppauksorn, Sunikom and Jildeh, Toufic R and Provencher, Matthew T},
title = {Unicortical button fixation provides higher strength compared with transosseous repair for subscapularis tendon in total shoulder arthroplasty},
journal = {Journal of Shoulder and Elbow Surgery},
year = {2025},
doi = {10.1016/j.jse.2024.07.019}
}
EUVIMED