High-resolution single-particle imaging at 100–200 keV with the Gatan Alpine direct electron detector
Journal of Structural Biology · 2024
DOI: 10.1016/j.jsb.2024.108108 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Structural Biology, 2024. Verfasst von Lieza M Chan und 13 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cellular and Molecular Pharmacology, University of California, San Francisco, CA 94158, United States. Hauptthemen laut MEDLINE: Cryoelectron Microscopy, Electrons, Apoferritins. Die Arbeit wurde bisher 17-mal zitiert (Crossref) und nennt 57 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIH HHS, NIGMS NIH HHS, Gatan.
Abstract (Originalsprache)
Developments in direct electron detector technology have played a pivotal role in enabling high-resolution structural studies by cryo-EM at 200 and 300 keV. Yet, theory and recent experiments indicate advantages to imaging at 100 keV, energies for which the current detectors have not been optimized. In this study, we evaluated the Gatan Alpine detector, designed for operation at 100 and 200 keV. Compared to the Gatan K3, Alpine demonstrated a significant DQE improvement at these energies, specifically a ∼ 4-fold improvement at Nyquist at 100 keV. In single-particle cryo-EM experiments, Alpine datasets yielded better than 2 Å resolution reconstructions of apoferritin at 120 and 200 keV on a ThermoFisher Scientific (TFS) Glacios microscope fitted with a non-standard SP-Twin lens. We also achieved a ∼ 3.2 Å resolution reconstruction of a 115 kDa asymmetric protein complex, proving Alpine's effectiveness with complex biological samples. In-depth analysis revealed that Alpine reconstructions are comparable to K3 reconstructions at 200 keV, and remarkably, reconstruction from Alpine at 120 keV on a TFS Glacios surpassed all but the 300 keV data from a TFS Titan Krios with GIF/K3. Additionally, we show Alpine's capability for high-resolution data acquisition and screening on lower-end systems by obtaining ∼ 3 Å resolution reconstructions of apoferritin and aldolase at 100 keV and detailed 2D averages of a 55 kDa sample using a side-entry cryo holder. Overall, we show that Gatan Alpine performs well with the standard 200 keV imaging systems and may potentially capture the benefits of lower accelerating voltages, bringing smaller sized particles within the scope of cryo-EM.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Cellular and Molecular Pharmacology, University of California, San Francisco, CA 94158, United States.
- Department of Integrative Structural and Computational Biology, Scripps Research, La Jolla, CA 92024, United States.
- Department of Biochemistry & Biophysics, University of California, San Francisco, CA 94158, United States.
- Gatan Inc., Pleasanton, CA, United States.
- Department of Integrative Structural and Computational Biology, Scripps Research, La Jolla, CA 92024, United States. Electronic address: glander@scripps.edu.
- Department of Cellular and Molecular Pharmacology, University of California, San Francisco, CA 94158, United States. Electronic address: kliment.verba@ucsf.edu.
Förderung
- NIH HHS
- NIGMS NIH HHS
- Gatan
- University of California, San Francisco
- George E. Hewitt Foundation for Medical Research
- National Institutes of Health
Zitieren
APA
Chan, L. M., Courteau, B. J., Maker, A., Wu, M., Basanta, B., Mehmood, H., Bulkley, D., Joyce, D., Lee, B. C., Mick, S., Czarnik, C., Gulati, S., Lander, G. C., & Verba, K. A. (2024). High-resolution single-particle imaging at 100–200 keV with the Gatan Alpine direct electron detector. Journal of Structural Biology. https://doi.org/10.1016/j.jsb.2024.108108
Vancouver
Chan LM, Courteau BJ, Maker A, Wu M, Basanta B, Mehmood H, et al. High-resolution single-particle imaging at 100–200 keV with the Gatan Alpine direct electron detector. Journal of Structural Biology. 2024. doi:10.1016/j.jsb.2024.108108
BibTeX
@article{Chan2024,
author = {Chan, Lieza M and Courteau, Brandon J and Maker, Allison and Wu, Mengyu and Basanta, Benjamin and Mehmood, Hev and Bulkley, David and Joyce, David and Lee, Brian C and Mick, Stephen and Czarnik, Cory and Gulati, Sahil and Lander, Gabriel C and Verba, Kliment A},
title = {High-resolution single-particle imaging at 100–200 keV with the Gatan Alpine direct electron detector},
journal = {Journal of Structural Biology},
year = {2024},
doi = {10.1016/j.jsb.2024.108108}
}
EUVIMED