Fachartikel in der Zeitschrift Journal of Pharmaceutical and Biomedical Analysis, 2024. Verfasst von Olajide Adetunji und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Chemistry, Center for Biomedical Research, Missouri University of Science and Technology, Rolla, MO 65409, USA. Hauptthemen laut MEDLINE: Solvents, Limit of Detection. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 21 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Pharmaceutical manufacturing utilizes solvents at different stages of production. Some of the harmful solvent residuals may be retained in the final product; therefore, they need to be monitored for quality control and to meet the regulation requirement. Here, a novel method capable of rapidly analyzing residual solvents in pharmaceutical products was developed using a compact-portable gas chromatography with a photoionization detector (GC-PID). The method consists of modified Tedlar® bag sampling, online pre-concentration, separation of volatiles by miniaturized GC, and micro-PID detection. The method detection limits of selected residual solvents were in the range of 26.00 - 52.03 pg/mL which is much lower than the pharmaceutical compliance concentration limits. Limits of detection > 520 pg of analyte per grams of sample was also determined for the over-the-counter drugs. The method performance showed rapid speed (5 min), linear calibration (r2 91.2 %) and precision (RSD < 6.5 %) for the selected residual solvents, including 1,4-dioxane, benzene, chlorobenzene, cyclohexane, xylenes, and toluene. This portable and rapid method could be deployed during the pharmaceutical drug manufacturing processes for quality control.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

SolventsLimit of DetectionChromatography, GasDrug ContaminationPharmaceutical PreparationsQuality ControlReproducibility of ResultsCalibration

Stichwörter der Autor:innen

On-line pre-concentrationPharmaceutical productsPortable GC with photoionization detectorVolatile residual solvents

Substanzen

SolventsPharmaceutical Preparations

Autor:innen und Institutionen

  • Department of Chemistry, Center for Biomedical Research, Missouri University of Science and Technology, Rolla, MO 65409, USA.
  • Department of Chemistry, Center for Biomedical Research, Missouri University of Science and Technology, Rolla, MO 65409, USA. Electronic address: nam@mst.edu.

Zitieren

APA

Adetunji, O., Kaur, S., Wheeler, N., Nam, V., Shi, H., & Nam, P. (2024). Monitoring residual solvents in pharmaceutical products using a portable pre-concentration GC-PID. Journal of Pharmaceutical and Biomedical Analysis. https://doi.org/10.1016/j.jpba.2024.116327

Vancouver

Adetunji O, Kaur S, Wheeler N, Nam V, Shi H, Nam P. Monitoring residual solvents in pharmaceutical products using a portable pre-concentration GC-PID. Journal of Pharmaceutical and Biomedical Analysis. 2024. doi:10.1016/j.jpba.2024.116327

BibTeX
@article{Adetunji2024,
  author = {Adetunji, Olajide and Kaur, Sargun and Wheeler, Nicole and Nam, Victor and Shi, Honglan and Nam, Paul},
  title = {Monitoring residual solvents in pharmaceutical products using a portable pre-concentration GC-PID},
  journal = {Journal of Pharmaceutical and Biomedical Analysis},
  year = {2024},
  doi = {10.1016/j.jpba.2024.116327}
}
Kostenlos starten →