Mechanical studies for the rotational stability of a cemented stem in cases with stem anteversion adjustment in the cement mantle
Journal of Orthopaedic Science · 2025
DOI: 10.1016/j.jos.2024.10.006 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Journal of Orthopaedic Science, 2025. Verfasst von Ariha Goshi und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopedic Surgery, Goshi Hospital, 10-8-20, Nagasunishidoori, Amagasaki City, Hyogo, 660-0807, Japan. Hauptthemen laut MEDLINE: Bone Cements, Hip Prosthesis, Cementation, Arthroplasty, Replacement, Hip. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 28 Literaturangaben.
Abstract (Originalsprache)
INTRODUCTION: The cemented stem can accommodate various shapes of femurs and is a stable stem with long-term results. Additionally, the cemented stem is a highly versatile stem that can be used to adjust stem anteversion in patients with variable femoral anteversion. On the other hand, rotating a stem with increasing cement viscosity may result in reduced postoperative rotational stability. MATERIALS AND METHODS: Rotational stability was compared using two different cementing techniques via in vitro models. Six stems in Group S (Static) were inserted and fixed until the cement was completely hardened. Six stems in Group R (Rotation) were inserted and then rotated 15° before the cement had fully cured. After the cement had completely set, a static rotation testing was performed in which the stem was rotated at a rate of 1°/sec to 20° while a vertical load of 600 N was applied. For the dynamic testing, a vertical load of 600 N and 10,000 cycles of repetitive rotation were applied at 10° around the stem axis at 0.1 Hz. After 10,000 cycles, maximum torque was compared between Groups S and R. RESULTS: In both groups, the torque value increased proportionally to the rotation angle. There were no significant differences in resistance to static rotation between Groups S and R (P = 0.599). In the dynamic testing, there was no significant difference between Groups S and R in both minimum torque and maximum torque in any of the cycles. There were no differences in the rotational torque between Groups S and R at 10,000 cycles (P = 0.693). CONCLUSIONS: The rotational stability of the cemented stems did not decrease in vitro when stem version control was performed up to 15° before the cement had completely hardened.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Orthopedic Surgery, Goshi Hospital, 10-8-20, Nagasunishidoori, Amagasaki City, Hyogo, 660-0807, Japan.
- Depertment of Orthopedic Surgery, Hyogo Medical University Hospital, 1-1, Mukogawa-cho, Nishinomiya City, Hyogo, 663-8501, Japan.
- Department of Orthopedic Surgery, Nishinomya Kaisei Hospital, 1-4, Ohama-cho, Nishinomiya City, Hyogo, 662-0957, Japan.
- Department of Orthopedic Surgery, Nishinomya Kaisei Hospital, 1-4, Ohama-cho, Nishinomiya City, Hyogo, 662-0957, Japan. Electronic address: fukunishi.shigeo@hmw.gr.jp.
Zitieren
APA
Goshi, A., Takeda, Y., Nakai, T., & Fukunishi, S. (2025). Mechanical studies for the rotational stability of a cemented stem in cases with stem anteversion adjustment in the cement mantle. Journal of Orthopaedic Science. https://doi.org/10.1016/j.jos.2024.10.006
Vancouver
Goshi A, Takeda Y, Nakai T, Fukunishi S. Mechanical studies for the rotational stability of a cemented stem in cases with stem anteversion adjustment in the cement mantle. Journal of Orthopaedic Science. 2025. doi:10.1016/j.jos.2024.10.006
BibTeX
@article{Goshi2025,
author = {Goshi, Ariha and Takeda, Yu and Nakai, Takuya and Fukunishi, Shigeo},
title = {Mechanical studies for the rotational stability of a cemented stem in cases with stem anteversion adjustment in the cement mantle},
journal = {Journal of Orthopaedic Science},
year = {2025},
doi = {10.1016/j.jos.2024.10.006}
}
EUVIMED