Optimal additional support screw position for prevention of hinge fracture in biplanar closed wedge distal femoral osteotomy

Masaru Higa, Hiroshi Nakayama, Ryo Kanto, Shintaro Onishi, Shinichi Yoshiya, Toshiya Tachibana, Tomoya Iseki

Journal of Orthopaedics · 2023-09 · 43 · 1-5

DOI: 10.1016/j.jor.2023.07.017 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Orthopaedics, 2023-09. Verfasst von Masaru Higa und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Mechanical Engineering, University of Hyogo, 2167 shosha, Himeji, Hyogo, 671-2280, Japan. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 24 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The purpose of this study was to examine the biomechanical significance of supplemental fixation using a positional screw in prevention of the hinge fracture in lateral closed-wedge distal femoral osteotomy (LCW-DFO) by means of a three-dimensional finite element analysis.

Methods

The three-dimensional numerical knee models with LCW-DFO were developed. To assess the mechanical efficacy of the positional screw and determine its optimal position and orientation, in total, 13 screwing methods were analyzed. In the first four methods, the screw was supported by the cortical bone only on the medial surface (mono-cortical). In the other 9 models, the screw was supported by both medial and lateral cortical bones (bi-cortical). Under 1000 N of vertical force and 5 Nm of rotational torques, the highest shear stress value around the medial hinge area was adopted as an analytical parameter.

Results

In mono-cortical methods, with the cancellous bone support, all methods were able to reduce the highest stress value compared to the value without the screw, while the efficacy was rather inferior when the screw was in horizontal direction. Without the cancellous bone support, however, all methods were not able to reduce the stress value. In bi-cortical methods, with the cancellous bone support, almost all screw augmentation methods were able to reduce the stress value. When screwing from the medial to the lateral, it only gets worse when going extremely posterior. Without the cancellous bone support, all methods were able to reduce the stress value.

Conclusion

The mechanical efficacy of the bi-cortical method was proven regardless of the quality of the local cancellous bone.

Schlagwörter

Stichwörter der Autor:innen

Distal femur osteotomyHinge fractureKneeOsteotomyScrew fixation

Autor:innen und Institutionen

  • Department of Mechanical Engineering, University of Hyogo, 2167 shosha, Himeji, Hyogo, 671-2280, Japan.
  • Department of Orthopaedic Surgery, Hyogo Medical University, Mukogawa-cho, Nishinomiya, Hyogo, 653-8501, Japan.
  • Department of Orthopaedic Surgery, Nishinomiya Kaisei Hospital, 1-4 Ohamacho, Nishinomiya, Hyogo, 662-0957, Japan.

Zitieren

APA

Higa, M., Nakayama, H., Kanto, R., Onishi, S., Yoshiya, S., Tachibana, T., & Iseki, T. (2023). Optimal additional support screw position for prevention of hinge fracture in biplanar closed wedge distal femoral osteotomy. Journal of Orthopaedics, 43, 1-5. https://doi.org/10.1016/j.jor.2023.07.017

Vancouver

Higa M, Nakayama H, Kanto R, Onishi S, Yoshiya S, Tachibana T, et al. Optimal additional support screw position for prevention of hinge fracture in biplanar closed wedge distal femoral osteotomy. Journal of Orthopaedics. 2023;43:1-5. doi:10.1016/j.jor.2023.07.017

BibTeX
@article{Higa2023,
  author = {Higa, Masaru and Nakayama, Hiroshi and Kanto, Ryo and Onishi, Shintaro and Yoshiya, Shinichi and Tachibana, Toshiya and Iseki, Tomoya},
  title = {Optimal additional support screw position for prevention of hinge fracture in biplanar closed wedge distal femoral osteotomy},
  journal = {Journal of Orthopaedics},
  year = {2023},
  volume = {43},
  pages = {1--5},
  doi = {10.1016/j.jor.2023.07.017}
}
Kostenlos starten →