Lumbar Spine Densitometry in People with Spinal Cord Injury: Investigation of Potential Sources of Errors
Journal of Clinical Densitometry · 2024
DOI: 10.1016/j.jocd.2024.101528 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Clinical Densitometry, 2024. Verfasst von Matteo Ponzano und 5 weiteren Autor:innen. Beteiligte Einrichtung: School of Health and Exercise Sciences, The University of British Columbia, Kelowna, BC Canada; International Collaboration on Repair Discoveries (IC…. Hauptthemen laut MEDLINE: Spinal Cord Injuries, Lumbar Vertebrae, Absorptiometry, Photon, Bone Density. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Craig H Neilsen Foundation, Michael Smith Health Research BC, Research Institute for Aging, University of Waterloo.
Abstract (Originalsprache)
PURPOSE: People with spinal cord injury (SCI) experience a considerable loss of bone after the injury. Lumbar spine (LS) bone mineral density (BMD) has been reported to be within the normal range, or even higher when assessed with DXA, in people with SCI; hence, it has been hypothesized that sources of error may spuriously increase LS BMD. The aim of this study was to describe the frequency of potential sources of error that may alter LS BMD measurement in a cohort of individuals with chronic SCI at baseline and over a 2-year period. METHODS: We analyzed baseline and 2-year follow up DXA scans (Hologic Discovery QDR 4500, Hologic Inc., MA, USA) previously performed from a cohort of males and females with chronic SCI. Two physicians independently reviewed each scan, commented on whether the scan was appropriate for BMD analysis, should be re-analyzed, or be removed from the dataset, and reported on the presence of potential sources of error in LS BMD measurement. RESULTS: We reviewed 115 lumbar spine DXA scans from 58 participants, and 107 (93.0 %) scans from 52 participants presented at least one potential source of error. At baseline, the average number of potential sources of error per scan was 5.5 ± 1.7 and 5.7 ± 1.5 according to rater 1 and rater 2, respectively. Follow-up scans presented an average of 5.6 ± 1.6 and 5.7 ± 1.4 potential sources of error according to rater 1 and rater 2, respectively. Facet sclerosis, osteophytes and difficulty in detecting bone edges were the most prevalent sources of error. CONCLUSION: The high frequency of potential sources of error is consistent with current recommendations against the use of LS BMD for fracture risk assessment in people with SCI.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- School of Health and Exercise Sciences, The University of British Columbia, Kelowna, BC Canada; International Collaboration on Repair Discoveries (ICORD), Blusson Spinal Cord Centre (BSCC), The University of British Col…
- Department of Kinesiology and Health Sciences, University of Waterloo, Waterloo, ON, Canada; Schlegel-UW Research Institute for Aging, University of Waterloo, Waterloo, ON, Canada; KITE Research Institute, University He…
- KITE Research Institute, University Health Network, Toronto, Canada; Rehabilitation Sciences Institute, Temerty Faculty of Medicine, University of Toronto, Toronto, Canada; Department of Medicine, Temerty Faculty of Med…
- KITE Research Institute, University Health Network, Toronto, Canada. Electronic address: sivakumar.gulasingam@uhn.ca.
- Department of Kinesiology and Health Sciences, University of Waterloo, Waterloo, ON, Canada. Electronic address: jack.callaghan@uwaterloo.ca.
- Department of Kinesiology and Health Sciences, University of Waterloo, Waterloo, ON, Canada; KITE Research Institute, University Health Network, Toronto, Canada; Rehabilitation Sciences Institute, Temerty Faculty of Med…
Förderung
- Craig H Neilsen Foundation
- Michael Smith Health Research BC
- Research Institute for Aging, University of Waterloo
- University of Toronto
- Natural Sciences and Engineering Research Council of Canada
Zitieren
APA
Ponzano, M., Giangregorio, L. M., Furlan, J. C., Gulasingam, S., Callaghan, J. P., & Craven, B. C. (2024). Lumbar Spine Densitometry in People with Spinal Cord Injury: Investigation of Potential Sources of Errors. Journal of Clinical Densitometry. https://doi.org/10.1016/j.jocd.2024.101528
Vancouver
Ponzano M, Giangregorio LM, Furlan JC, Gulasingam S, Callaghan JP, Craven BC. Lumbar Spine Densitometry in People with Spinal Cord Injury: Investigation of Potential Sources of Errors. Journal of Clinical Densitometry. 2024. doi:10.1016/j.jocd.2024.101528
BibTeX
@article{Ponzano2024,
author = {Ponzano, Matteo and Giangregorio, Lora M and Furlan, Julio C and Gulasingam, Sivakumar and Callaghan, Jack P and Craven, B Catharine},
title = {Lumbar Spine Densitometry in People with Spinal Cord Injury: Investigation of Potential Sources of Errors},
journal = {Journal of Clinical Densitometry},
year = {2024},
doi = {10.1016/j.jocd.2024.101528}
}
EUVIMED