Sensor-based measures of knee brace adherence have low agreement with self-report methods: A multi-measure study among knee osteoarthritis patients
Osteoarthritis and Cartilage · 2026
DOI: 10.1016/j.joca.2026.06.005 ↗Kurz gefasst
Randomisierte kontrollierte Studie in der Zeitschrift Osteoarthritis and Cartilage, 2026. Verfasst von Michael J Callaghan und 11 weiteren Autor:innen. Beteiligte Einrichtung: Department of Health Professions, Faculty of Health and Education, Manchester Metropolitan University, Manchester, UK; NIHR Manchester Biomedical Res…. Hauptthemen laut MEDLINE: Osteoarthritis, Knee, Braces, Patient Compliance, Self Report. Die Arbeit nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Institute for Health Research.
Abstract (Originalsprache)
OBJECTIVE: To explore agreement between self-report and objectively measured adherence to brace wearing by patients with knee osteoarthritis. METHOD: A single-arm observational analysis nested within the PROP OA randomised controlled trial (ISRCTN28555470). Of 237 adults with symptomatic knee osteoarthritis randomised to brace treatment, 60 were included in this sub-study investigating three different methods of assessing knee brace wear time over 26 weeks: 1. Self-report questionnaires (SRQ) at 12 weeks and 26 weeks; 2. Short message service (SMS) questions (days worn in past week, typical hours per day when worn) administered from week 1 to week 24; 3. A skin temperature sensor embedded in the brace, sampling every 10 min for 26 weeks. The presence and reason for the sensor were concealed from participants. The estimated proportion of participants meeting "minimum brace use", defined a priori as ≥1 h on ≥2 days in past week, was described for each measurement method, overall and by brace type (unloader, neutral). For temperature sensor measurements, time spent above 24°C and time spent above 25°C were used. Agreement between the measures was summarised by percentage agreement and kappa (ĸ). RESULTS: The estimated proportions of participants meeting "minimum brace use" at 12 weeks were 83% (SRQ), 83% (SMS), 60% and 58% (temperature sensor, 24°C and 25°C thresholds, respectively). At 26 weeks, the corresponding estimates reduced to 72%, 71% (SMS at 24 weeks), 43% and 37%. Sensor data suggested the sharpest decline in brace use occurred within the first 12 weeks. Agreement between self-report measures was higher than between self-report measures and sensor (SRQ vs SMS at 12 weeks: 92% agreement, ĸ=0.67 (95%CI: 0.34, 1.00); SRQ vs Sensor at 12 weeks: 74%, 0.35 (0.10, 0.60); SMS vs Sens at 12 weeks: 76%, 0.36 (0.05, 0.66). Agreement between all measurement methods reduced at 26 weeks. CONCLUSIONS: This novel use of a temperature sensor to monitor brace adherence in knee osteoarthritis indicates that self-report adherence substantially overestimates knee brace wearing time, with implications for clinical trials and practice.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Health Professions, Faculty of Health and Education, Manchester Metropolitan University, Manchester, UK; NIHR Manchester Biomedical Research Centre, Manchester University Foundation NHS Trust, Manchester,…
- NIHR Manchester Biomedical Research Centre, Manchester University Foundation NHS Trust, Manchester, UK; School of Medicine, Boston University, Boston, MA, USA.
- School of Medicine, Primary Care Centre Versus Arthritis, Keele University, Keele, UK.
- Faculty of Biology, Medicine & Health, The University of Manchester, Manchester, UK; Henry M. Goldman School of Dental Medicine, Boston University, Boston, Mass, USA; School of Health Sciences, Division of Psychology an…
- School of Medicine, Primary Care Centre Versus Arthritis, Keele University, Keele, UK; Keele Clinical Trials Unit, Keele University, Keele, UK.
- Medical Research Council Versus Arthritis Centre for Integrated Research into Musculoskeletal Ageing, Population & Health Sciences Institute, Newcastle University, Newcastle upon Tyne, Tyne and Wear, UK; Northumbria Hea…
Förderung
- National Institute for Health Research
Zitieren
APA
Callaghan, M. J., Felson, D. T., Holden, M. A., Borrelli, B., Nicholls, E., Birrell, F., Foster, N. E., Halliday, N., Ingram, C., Jinks, C., Peat, G. M., & PROP OA team (see table S1) (2026). Sensor-based measures of knee brace adherence have low agreement with self-report methods: A multi-measure study among knee osteoarthritis patients. Osteoarthritis and Cartilage. https://doi.org/10.1016/j.joca.2026.06.005
Vancouver
Callaghan MJ, Felson DT, Holden MA, Borrelli B, Nicholls E, Birrell F, et al. Sensor-based measures of knee brace adherence have low agreement with self-report methods: A multi-measure study among knee osteoarthritis patients. Osteoarthritis and Cartilage. 2026. doi:10.1016/j.joca.2026.06.005
BibTeX
@article{Callaghan2026,
author = {Callaghan, Michael J and Felson, David T and Holden, Melanie A and Borrelli, Belinda and Nicholls, Elaine and Birrell, Fraser and Foster, Nadine E and Halliday, Nicola and Ingram, Carol and Jinks, Clare and Peat, George M and PROP OA team (see table S1), PROP OA team (see table S1)},
title = {Sensor-based measures of knee brace adherence have low agreement with self-report methods: A multi-measure study among knee osteoarthritis patients},
journal = {Osteoarthritis and Cartilage},
year = {2026},
doi = {10.1016/j.joca.2026.06.005}
}
EUVIMED