EUVIMEDEUVIMED entdecken
Open Access

Precisely-timed outpatient recordings of subcortical local field potentials from wireless streaming-capable deep-brain stimulators: a method and toolbox

Cheol Soh, Mario Hervault, Andrea H Rohl, Jeremy D W Greenlee, Jan R Wessel

Journal of Neuroscience Methods · 2025-06 · 418 · 110448

DOI: 10.1016/j.jneumeth.2025.110448 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Neuroscience Methods, 2025-06. Verfasst von Cheol Soh und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Psychological and Brain Sciences, University of Iowa, United States; Cognitive Control Collaborative, University of Iowa, United States…. Hauptthemen laut MEDLINE: Deep Brain Stimulation, Wireless Technology, Electroencephalography, Brain. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 53 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NINDS NIH HHS, National Science Foundation, National Institutes of Health.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Investigations of the electrophysiological mechanisms of the human subcortex have relied on recording local field potentials (LFPs) during deep-brain stimulation (DBS) neurosurgery. However, the neurosurgical setting severely restricts the research use of these recordings. Recently developed sensing-capable DBS devices wirelessly stream subcortical LFPs in outpatient settings. These recordings have tremendous potential for research. However, synchronizing them with other behavior or neural recordings is challenging, as the clinical devices do not accept digital timing information. NEW METHOD: Switching the DBS device on introduces transient yet consistent artifacts in both the LFP and simultaneous scalp-EEG recordings. We use these artifacts as a reference to align these recordings (N = 20). We tested whether the alignment was precise enough to match a ground truth state (large artifacts produced by transcranial magnetic stimulation, TMS), yielded trial-averaged event-locked LFPs, and phase consistency across trials. We further evaluated the consistency of task-related LFPs across outpatient and perisurgical recordings. RESULTS AND COMPARISON WITH EXISTING METHOD(S): Previous alignment methods were limited because they relied on inconsistent on/offset features of DBS artifacts caused by ongoing stimulation. Moreover, they only provided limited validation. Our highly precise alignment method showed a maximum deviation of only 8 ms - clearly superior to prior techniques. Furthermore, event-related activity patterns were comparable across outpatient and perisurgical LFP recordings. CONCLUSIONS: We present a method and a MATLAB toolbox that inserts the most precise digital timing information into wirelessly-streamed DBS-LFP recordings to date. By enabling event-related research with high-temporal precision, this method greatly enhances the utility of these recordings.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Deep Brain StimulationWireless TechnologyElectroencephalographyBrainHumansMaleFemaleArtifactsAdultOutpatientsMiddle AgedSignal Processing, Computer-Assisted

Stichwörter der Autor:innen

Deep brain stimulationEEG/LFP alignmentLocal field potentialsPercept DBSToolbox

Autor:innen und Institutionen

  • Department of Psychological and Brain Sciences, University of Iowa, United States; Cognitive Control Collaborative, University of Iowa, United States. Electronic address: cheol-soh@uiowa.edu.
  • Department of Psychological and Brain Sciences, University of Iowa, United States; Cognitive Control Collaborative, University of Iowa, United States.
  • Department of Neurosurgery, University of Iowa, United States.
  • Department of Psychological and Brain Sciences, University of Iowa, United States; Cognitive Control Collaborative, University of Iowa, United States; Department of Neurology, University of Iowa, United States.

Förderung

  • NINDS NIH HHS
  • National Science Foundation
  • National Institutes of Health
  • National Institute of Neurological Disorders and Stroke

Zitieren

APA

Soh, C., Hervault, M., Rohl, A. H., Greenlee, J. D. W., & Wessel, J. R. (2025). Precisely-timed outpatient recordings of subcortical local field potentials from wireless streaming-capable deep-brain stimulators: a method and toolbox. Journal of Neuroscience Methods, 418, 110448. https://doi.org/10.1016/j.jneumeth.2025.110448

Vancouver

Soh C, Hervault M, Rohl AH, Greenlee JDW, Wessel JR. Precisely-timed outpatient recordings of subcortical local field potentials from wireless streaming-capable deep-brain stimulators: a method and toolbox. Journal of Neuroscience Methods. 2025;418:110448. doi:10.1016/j.jneumeth.2025.110448

BibTeX
@article{Soh2025,
  author = {Soh, Cheol and Hervault, Mario and Rohl, Andrea H and Greenlee, Jeremy D W and Wessel, Jan R},
  title = {Precisely-timed outpatient recordings of subcortical local field potentials from wireless streaming-capable deep-brain stimulators: a method and toolbox},
  journal = {Journal of Neuroscience Methods},
  year = {2025},
  volume = {418},
  pages = {110448},
  doi = {10.1016/j.jneumeth.2025.110448}
}
Kostenlos starten →