Interrater and Intrarater Reliability of a Handheld Myotonometer in Measuring Mechanical Properties of the Neck and Orofacial Muscles
Journal of Manipulative and Physiological Therapeutics · 2021-01 · 44 (1) · 42-48
DOI: 10.1016/j.jmpt.2020.08.002 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift Journal of Manipulative and Physiological Therapeutics, 2021-01. Verfasst von Serkan Taş und 2 weiteren Autor:innen. Beteiligte Einrichtung: Toros University, School of Health Sciences, Department of Physiotherapy and Rehabilitation, Mersin, Turkey. Electronic address: serkntas@gmail.com. Die Arbeit wurde bisher 48-mal zitiert (Crossref) und nennt 22 Literaturangaben.
Abstract (Originalsprache)
OBJECTIVE: The purpose this study was to investigate the reliability of a handheld myotonometer in measuring the mechanical properties of the neck and orofacial muscles in asymptomatic individuals. METHODS: The study included 16 healthy participants. The mechanical properties (frequency, decrement, stiffness, relaxation time, and creep) of the selected muscles were measured with a MyotonPRO myotonometer (Mumeetria Ltd, Tallinn, Estonia). The sternocleidomastoid, upper trapezius, cervical extensor, and masseter muscles were selected to determine the reliability of the device. Measurements were performed by 2 examiners to determine interrater reliability; for intrarater reliability, an examiner repeated the measurements 1 week after the first measurements. RESULTS: The results revealed moderate to excellent intrarater and interrater reliability (intraclass correlation coefficients: 0.50-0.95) in measuring muscle mechanic properties. The standard error of measurement in the tested muscles ranged from 0.3 to 0.8 Hz for frequency, from 7.4 to 20.9 N/m for stiffness, from 0.1 to 0.2 for decrement, and from 0.8 to 1.4 ms for relaxation time. The minimum detectable change ranged from 0.8 to 2.2 Hz for frequency, from 20.5 to 57.9 N/m for stiffness, from 0.2 to 0.6 for decrement, from 2.2 to 3.9 ms for relaxation time, and from 0.2 to 0.3 for creep. In addition, the coefficients of variation were below 9.1% for all the assessed parameters. CONCLUSION: The obtained results demonstrate that the MyotonPRO device is a reliable and repeatable tool to quantify the frequency, stiffness, decrement, relation time, and creep of the neck and orofacial muscles in asymptomatic individuals.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Toros University, School of Health Sciences, Department of Physiotherapy and Rehabilitation, Mersin, Turkey. Electronic address: serkntas@gmail.com.
- Ardahan University, Nihat Delibalta Vocational School, Laboratory and Veterinary Health Department, Ardahan, Turkey.
- Toros University, School of Health Sciences, Department of Physiotherapy and Rehabilitation, Mersin, Turkey.
Zitieren
APA
Taş, S., Yaşar, Ü., & Kaynak, B. A. (2021). Interrater and Intrarater Reliability of a Handheld Myotonometer in Measuring Mechanical Properties of the Neck and Orofacial Muscles. Journal of Manipulative and Physiological Therapeutics, 44(1), 42-48. https://doi.org/10.1016/j.jmpt.2020.08.002
Vancouver
Taş S, Yaşar Ü, Kaynak BA. Interrater and Intrarater Reliability of a Handheld Myotonometer in Measuring Mechanical Properties of the Neck and Orofacial Muscles. Journal of Manipulative and Physiological Therapeutics. 2021;44(1):42-48. doi:10.1016/j.jmpt.2020.08.002
BibTeX
@article{Ta2021,
author = {Taş, Serkan and Yaşar, Ümit and Kaynak, Besime Ahu},
title = {Interrater and Intrarater Reliability of a Handheld Myotonometer in Measuring Mechanical Properties of the Neck and Orofacial Muscles},
journal = {Journal of Manipulative and Physiological Therapeutics},
year = {2021},
volume = {44},
number = {1},
pages = {42--48},
doi = {10.1016/j.jmpt.2020.08.002}
}