Fachartikel in der Zeitschrift Journal of the Mechanical Behavior of Biomedical Materials, 2025-05. Verfasst von Hüray Ilayda Kök und 7 weiteren Autor:innen. Beteiligte Einrichtung: Leibniz University Hannover, Institute of Continuum Mechanics, Hannover, Germany. Electronic address: koek@ikm.uni-hannover.de. Hauptthemen laut MEDLINE: Dental Implants, Stress, Mechanical, Dental Prosthesis Design. Die Arbeit wurde bisher 33-mal zitiert (Crossref) und nennt 72 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Deutsche Forschungsgemeinschaft.
Abstract (Originalsprache)
To improve longevity and performance of dental implants by reducing stress-shielding, a modification of the internal structure of the implant can be a solution. In this paper the inner design of the implant is generated either using a topology optimization approach or an approach based on TPMS lattice structures. These approaches aim to maintain long-term stability and to reduce stress-shielding. For both approaches, the mechanostat model was applied to investigate the influence of the inner structure to the surrounding bone tissue and compare the standard uniform implant. For the investigation an ANSYS model was used with material parameters obtained from a mechanical test of additively manufactured Ti6Al4V. Compared to the uniform implant, the topology-optimized implant showed 20% less stress-shielding, and the implant with triply periodic minimal surface structures (TPMS) showed 15% less stress-shielding. Further, the long-term-stability was investigated by introducing a high-cycle fatigue material model. Despite a change in the internal structure and a 45% reduction in the mass of the topology-optimized implant, the cycle numbers specified in the DIN EN ISO 14801 standard were fulfilled.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Leibniz University Hannover, Institute of Continuum Mechanics, Hannover, Germany. Electronic address: koek@ikm.uni-hannover.de.
- Leibniz University Hannover, Institute of Continuum Mechanics, Hannover, Germany.
- Department of Prosthetic Dentistry and Biomedical Materials Science, Hannover Medical School, Hannover, Germany.
- Chair of Materials Test Engineering (WPT), TU Dortmund University, Dortmund, Germany.
Förderung
- Deutsche Forschungsgemeinschaft
Zitieren
APA
Kök, H. I., Kick, M., Akbas, O., Stammkötter, S., Greuling, A., Stiesch, M., Walther, F., & Junker, P. (2025). Reduction of stress-shielding and fatigue-resistant dental implant design through topology optimization and TPMS lattices. Journal of the Mechanical Behavior of Biomedical Materials, 165, 106923. https://doi.org/10.1016/j.jmbbm.2025.106923
Vancouver
Kök HI, Kick M, Akbas O, Stammkötter S, Greuling A, Stiesch M, et al. Reduction of stress-shielding and fatigue-resistant dental implant design through topology optimization and TPMS lattices. Journal of the Mechanical Behavior of Biomedical Materials. 2025;165:106923. doi:10.1016/j.jmbbm.2025.106923
BibTeX
@article{Kk2025,
author = {Kök, Hüray Ilayda and Kick, Miriam and Akbas, Osman and Stammkötter, Sebastian and Greuling, Andreas and Stiesch, Meike and Walther, Frank and Junker, Philipp},
title = {Reduction of stress-shielding and fatigue-resistant dental implant design through topology optimization and TPMS lattices},
journal = {Journal of the Mechanical Behavior of Biomedical Materials},
year = {2025},
volume = {165},
pages = {106923},
doi = {10.1016/j.jmbbm.2025.106923}
}
EUVIMED