EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC-ND 4.0

In vitro investigation of a new thrombus aspiration and autologous blood reinfusion system

Weixiao Li, Qiang Li, Shuiting Zhai, Tianxiao Li, Nick Cheshire, Zhidong Zhang, Kai Liang

Journal of Interventional Medicine · 2019

DOI: 10.1016/j.jimed.2019.05.004 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Interventional Medicine, 2019. Verfasst von Weixiao Li und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Vascular and Endovascular Surgery, Henan Provincial People's Hospital, Zhengzhou University People's Hospital, China. Die Arbeit nennt 18 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: The aim of the study was to evaluate the feasibility of a new venous-thrombus aspiration and autologous blood (auto-blood) reinfusion system. MATERIALS AND METHODS: We constructed the venous model from polyvinyl chloride (PVC) tubes and three-way unions using a fresh clot of chicken blood as the venous thrombus. Eight French and 12F aspiration catheters were used to aspirate the thrombus in the right-pulmonary-artery model, 8 French and 14F aspiration catheters were used in the inferior-vena cava model, and 8 French and 10F aspiration catheters were used in the left-iliofemoral-vein model. A thrombus filtration and auto-blood reinfusion bottle was used to filter the thrombus and re-infuse auto-blood. We evaluated the thrombus aspiration capability of each catheter by comparing pre-aspirated with the post-aspirated thrombus volume, and we evaluated the difference in aspiration capability between the two catheters in each model by comparing their thrombus aspiration rates. We used Student's t-test for statistical analysis. RESULTS: Differences between pre-aspirated and post-aspirated thrombus volumes for each catheter were insignificant, as were those between the thrombus aspiration rates of the two catheters in each venous model. Using the thrombus aspiration and auto-blood reinfusion system, each aspiration catheter could fluently aspirate the thrombus out of the venous model. CONCLUSION: In this study, we designed a new venous-thrombus aspiration system. This system could be used to aspirate acute venous thrombi and re-infuse autologous blood.

Schlagwörter

Stichwörter der Autor:innen

Aspiration catheterAutologous blood reinfusionDeep-venous thrombosisManual aspiration thrombectomyPercutaneous mechanical thrombectomy

Autor:innen und Institutionen

  • Department of Vascular and Endovascular Surgery, Henan Provincial People's Hospital, Zhengzhou University People's Hospital, China.
  • Department of Medical Imaging, Henan Provincial People's Hospital, Zhengzhou University People's Hospital, China.
  • Department of Interventional Radiology, Henan Provincial People's Hospital, Zhengzhou University People's Hospital, China.
  • Department of Vascular Surgery, Royal Brompton & Harefield NHS Foundation Trust, Imperial College London, China.

Zitieren

APA

Li, W., Li, Q., Zhai, S., Li, T., Cheshire, N., Zhang, Z., & Liang, K. (2019). In vitro investigation of a new thrombus aspiration and autologous blood reinfusion system. Journal of Interventional Medicine. https://doi.org/10.1016/j.jimed.2019.05.004

Vancouver

Li W, Li Q, Zhai S, Li T, Cheshire N, Zhang Z, et al. In vitro investigation of a new thrombus aspiration and autologous blood reinfusion system. Journal of Interventional Medicine. 2019. doi:10.1016/j.jimed.2019.05.004

BibTeX
@article{Li2019,
  author = {Li, Weixiao and Li, Qiang and Zhai, Shuiting and Li, Tianxiao and Cheshire, Nick and Zhang, Zhidong and Liang, Kai},
  title = {In vitro investigation of a new thrombus aspiration and autologous blood reinfusion system},
  journal = {Journal of Interventional Medicine},
  year = {2019},
  doi = {10.1016/j.jimed.2019.05.004}
}