Melting curve analysis reveals false-positive norovirus detection in a molecular syndromic panel
Journal of Clinical Virology · 2024-08 · 173 · 105697
DOI: 10.1016/j.jcv.2024.105697 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Clinical Virology, 2024-08. Verfasst von Nancy Matic und 11 weiteren Autor:innen. Beteiligte Einrichtung: Division of Medical Microbiology and Virology, St. Paul's Hospital, Vancouver, Canada; Department of Pathology and Laboratory Medicine, University Br…. Hauptthemen laut MEDLINE: Norovirus, Caliciviridae Infections, Feces, Gastroenteritis. Die Arbeit wurde bisher 21-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
BACKGROUND: Molecular syndromic panels can improve rapidity of results and ease clinical laboratory workflow, although caution has been raised for potential false-positive results. Upon implementation of a new panel for infectious diarrhea (BioFire® FilmArray® Gastrointestinal [GI] Panel, bioMérieux) in our clinical laboratory, a higher than expected number of stool samples with norovirus were detected. OBJECTIVES: The goal of this study was to investigate positive percent agreement and the false-positive rate of norovirus detected by the multiplex BioFire GI panel compared to a singleplex commercial assay. STUDY DESIGN: From October 2023 to January 2024, all prospective stool samples with a positive norovirus result by BioFire had melting curves reviewed manually using the BioFire FilmArray Torch System. Stool samples further underwent testing by a supplementary real-time RT-PCR assay (Xpert® Norovirus, Cepheid) for comparative analysis. RESULTS: Of the 50 stool samples with norovirus detected by BioFire, 18 (36 %) tested negative by Xpert (deemed "false-positives"). Furthermore, melting curve analysis revealed nearly all of these samples had atypical melting curve morphologies for the "Noro-1" target on BioFire (16/18, 89 %), which was statistically significant (Odds Ratio 173.2, 95 % CI [22.2, 5326.9], p < 0.0001). Stool samples with multiple pathogens detected by BioFire including norovirus were not more likely to produce false-positive norovirus results (Odds Ratio 1, 95 % CI [0.3, 3.3], p = 1). CONCLUSIONS: Although not described in the manufacturer's Instructions for Use, we propose routine manual review of melting curves for the BioFire GI panel prior to reporting, to mitigate potential false-positive norovirus results.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Division of Medical Microbiology and Virology, St. Paul's Hospital, Vancouver, Canada; Department of Pathology and Laboratory Medicine, University British Columbia, Vancouver, Canada. Electronic address: nmatic@mail.ubc…
- Division of Medical Microbiology and Virology, St. Paul's Hospital, Vancouver, Canada.
- Division of Medical Microbiology and Virology, St. Paul's Hospital, Vancouver, Canada; Department of Pathology and Laboratory Medicine, University British Columbia, Vancouver, Canada.
Zitieren
APA
Matic, N., Lawson, T., Young, M., Jang, W., Bilawka, J., Gowland, L., Ritchie, G., Leung, V., Payne, M., Stefanovic, A., Romney, M. G., & Lowe, C. F. (2024). Melting curve analysis reveals false-positive norovirus detection in a molecular syndromic panel. Journal of Clinical Virology, 173, 105697. https://doi.org/10.1016/j.jcv.2024.105697
Vancouver
Matic N, Lawson T, Young M, Jang W, Bilawka J, Gowland L, et al. Melting curve analysis reveals false-positive norovirus detection in a molecular syndromic panel. Journal of Clinical Virology. 2024;173:105697. doi:10.1016/j.jcv.2024.105697
BibTeX
@article{Matic2024,
author = {Matic, Nancy and Lawson, Tanya and Young, Matthew and Jang, Willson and Bilawka, Jennifer and Gowland, Leah and Ritchie, Gordon and Leung, Victor and Payne, Michael and Stefanovic, Aleksandra and Romney, Marc G and Lowe, Christopher F},
title = {Melting curve analysis reveals false-positive norovirus detection in a molecular syndromic panel},
journal = {Journal of Clinical Virology},
year = {2024},
volume = {173},
pages = {105697},
doi = {10.1016/j.jcv.2024.105697}
}
EUVIMED