Contemporary 1-Year Outcomes of Mitral Valve-in-Ring With Balloon-Expandable Aortic Transcatheter Valves in the U.S.

Mayra E Guerrero, Vinayak N Bapat, Paul Mahoney, Amar Krishnaswamy, Mackram F Eleid, Marvin H Eng, Pradeep Yadav, Megan Coylewright, Raj Makkar, Molly Szerlip, Tamim Nazif, Susheel Kodali und 8 weitere

JACC: Cardiovascular Interventions · 2024-04 · 17 (7) · 874-886

DOI: 10.1016/j.jcin.2024.02.012 ↗

Kurz gefasst

Fachartikel in der Zeitschrift JACC: Cardiovascular Interventions, 2024-04. Verfasst von Mayra E Guerrero und 19 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiovascular Medicine, Mayo Clinic, Rochester, Minnesota, USA. Electronic address: guerrero.mayra@mayo.edu. Hauptthemen laut MEDLINE: Heart Valve Prosthesis Implantation, Heart Valve Prosthesis, Mitral Valve Insufficiency. Die Arbeit wurde bisher 18-mal zitiert (Crossref) und nennt 16 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Edwards Lifesciences.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Adequate valve performance after surgical mitral valve repair with an annuloplasty ring is not always sustained over time. The risk of repeat mitral valve surgery may be high in these patients. Transcatheter mitral valve-in-ring (MViR) is emerging as an alternative for high-risk patients. OBJECTIVES: The authors sought to assess contemporary outcomes of MViR using third-generation balloon-expandable aortic transcatheter heart valves. METHODS: Patients who underwent MViR and were enrolled in the STDS/ACC TVT (Society of Thoracic Surgeons/American College of Cardiology Transcatheter Valve Therapy) Registry between August 2015 and December 2022 were analyzed. RESULTS: A total of 820 patients underwent MViR at 236 sites, mean age was 72.2 ± 10.4 years, 50.9% were female, mean STS score was 8.2% ± 6.9%, and most (78%) were in NYHA functional class III to IV. Mean left ventricular ejection fraction was 47.8% ± 14.2%, mean mitral gradient was 8.9 ± 7.0 mm Hg, and 75.5% had ≥ moderate mitral regurgitation. Access was transseptal in 93.9% with 88% technical success. All-cause mortality at 30 days was 8.3%, and at 1 year, 22.4%, with a reintervention rate of 9.1%. At 1-year follow-up, 75.6% were NYHA functional class I to II, Kansas City Cardiomyopathy Questionnaire score increased by 25.9 ± 29.1 points, mean mitral valve gradient was 8.4 ± 3.4 mm Hg, and 91.7% had ≤ mild mitral regurgitation. CONCLUSIONS: MViR with third-generation balloon-expandable aortic transcatheter heart valves is associated with a significant reduction in mitral regurgitation and improvement in symptoms at 1 year, but with elevated valvular gradients and a high reintervention rate. MViR is a reasonable alternative for high-risk patients unable undergo surgery who have appropriate anatomy for the procedure. (STS/ACC TVT Registry Mitral Module [TMVR]; NCT02245763).

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Heart Valve Prosthesis ImplantationHeart Valve ProsthesisMitral Valve InsufficiencyHumansFemaleMiddle AgedAgedAged, 80 and overMaleMitral ValveStroke VolumeTreatment OutcomeVentricular Function, LeftCardiac Catheterization

Stichwörter der Autor:innen

degenerated mitral bioprosthesesmitral regurgitationtransapicaltranscatheter mitral valve replacementtransseptalvalve-in-ring

Autor:innen und Institutionen

  • Department of Cardiovascular Medicine, Mayo Clinic, Rochester, Minnesota, USA. Electronic address: guerrero.mayra@mayo.edu.
  • Department of Cardiothoracic Surgery, Allina Health Minneapolis Heart Institute at Abbott Northwestern Hospital, Minneapolis, Minnesota, USA.
  • Division of Cardiology, Department of Cardiovascular Services, Brody School of Medicine, East Carolina University, Greenville, North Carolina, USA.
  • Cleveland Clinic Foundation, Cleveland, Ohio, USA.
  • Department of Cardiovascular Medicine, Mayo Clinic, Rochester, Minnesota, USA.
  • Division of Cardiology, Banner University Medical Center, Phoenix, Arizona, USA.

Förderung

  • Edwards Lifesciences

Zitieren

APA

Guerrero, M. E., Bapat, V. N., Mahoney, P., Krishnaswamy, A., Eleid, M. F., Eng, M. H., Yadav, P., Coylewright, M., Makkar, R., Szerlip, M., Nazif, T., Kodali, S., George, I., Greenbaum, A., Babaliaros, V., Kapadia, S., Rihal, C. S., Whisenant, B., Thourani, V. H., & McCabe, J. M. (2024). Contemporary 1-Year Outcomes of Mitral Valve-in-Ring With Balloon-Expandable Aortic Transcatheter Valves in the U.S. JACC: Cardiovascular Interventions, 17(7), 874-886. https://doi.org/10.1016/j.jcin.2024.02.012

Vancouver

Guerrero ME, Bapat VN, Mahoney P, Krishnaswamy A, Eleid MF, Eng MH, et al. Contemporary 1-Year Outcomes of Mitral Valve-in-Ring With Balloon-Expandable Aortic Transcatheter Valves in the U.S. JACC: Cardiovascular Interventions. 2024;17(7):874-886. doi:10.1016/j.jcin.2024.02.012

BibTeX
@article{Guerrero2024,
  author = {Guerrero, Mayra E and Bapat, Vinayak N and Mahoney, Paul and Krishnaswamy, Amar and Eleid, Mackram F and Eng, Marvin H and Yadav, Pradeep and Coylewright, Megan and Makkar, Raj and Szerlip, Molly and Nazif, Tamim and Kodali, Susheel and George, Isaac and Greenbaum, Adam and Babaliaros, Vasilis and Kapadia, Samir and Rihal, Charanjit S and Whisenant, Brian and Thourani, Vinod H and McCabe, James M},
  title = {Contemporary 1-Year Outcomes of Mitral Valve-in-Ring With Balloon-Expandable Aortic Transcatheter Valves in the U.S},
  journal = {JACC: Cardiovascular Interventions},
  year = {2024},
  volume = {17},
  number = {7},
  pages = {874--886},
  doi = {10.1016/j.jcin.2024.02.012}
}
Kostenlos starten →