LC-MS/MS and EMIT measure the whole blood concentration of cyclosporine A: The two methods yield concordant results within the dynamic range of the latter, but the former shows broader application scenarios
Journal of Chromatography B · 2024
DOI: 10.1016/j.jchromb.2024.124154 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Chromatography B, 2024. Verfasst von Yuan-Yuan Zhang und 9 weiteren Autor:innen. Beteiligte Einrichtung: Pharmaceutical Sciences Research Center, Department of Pharmacy, Children's Hospital of Nanjing Medical University, Nanjing, China. Hauptthemen laut MEDLINE: Cyclosporine, Tandem Mass Spectrometry, Drug Monitoring. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 34 Literaturangaben. Gefördert von: Jiangsu Pharmaceutical Association.
Abstract (Originalsprache)
Cyclosporine A (CsA) is a widely used immunosuppressive drug with a narrow therapeutic index and large individual differences. Its therapeutic and toxic effects are closely related to blood drug concentrations, requiring routine therapeutic drug monitoring (TDM). The current main methods for TDM of CsA are enzyme multiplied immunoassay technique (EMIT) and liquid chromatography-tandem mass spectrometry (LC-MS/MS). However, few study on the method comparison of the EMIT and LC-MS/MS for the measurement of whole blood CsA concentration in children has been reported. In this study, we developed a simple and sensitive LC-MS/MS assay for the determination of CsA, and 657 cases of CsA concentrations were determined from 197 pediatric patients by a routine EMIT assay and by the validated in-house LC-MS/MS method on the same batch of samples, aimed to address the aforementioned concern. Consistency between the two assays was evaluated using linear regression and Bland-Altman analysis. The linear range of LC-MS/MS was 0.500-2000 ng/mL and that of the EMIT was 40-500 ng/mL, respectively. Overall, the correlation between the two methods was significant (r-value ranging from 0.8842 to 0.9441). Unsatisfactory consistency was observed in the concentrations 0.05). In conclusion, the two methods were closely correlated, but the CsA concentration by LC-MS/MS assay was slightly higher than that by EMIT method. Switching from the EMIT assay to the LC-MS/MS method was acceptable, and the LC-MS/MS method will receive broader application in clinical settings due to its better analytical capabilities, but the results need to be further verified in different laboratories.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Pharmaceutical Sciences Research Center, Department of Pharmacy, Children's Hospital of Nanjing Medical University, Nanjing, China.
- School of Basic Medicine and Clinical Pharmacy, China Pharmaceutical University, Nanjing, China.
- Pharmaceutical Sciences Research Center, Department of Pharmacy, Children's Hospital of Nanjing Medical University, Nanjing, China. Electronic address: huyahui324@163.com.
- Pharmaceutical Sciences Research Center, Department of Pharmacy, Children's Hospital of Nanjing Medical University, Nanjing, China. Electronic address: cy.chen508@gmail.com.
Förderung
- Jiangsu Pharmaceutical Association
Zitieren
APA
Zhang, Y. Y., Guo, H. L., Wang, J., Wang, W. J., Li, Y., Chu, C. C., Wu, C. Y., Huang, J., Hu, Y. H., & Chen, F. (2024). LC-MS/MS and EMIT measure the whole blood concentration of cyclosporine A: The two methods yield concordant results within the dynamic range of the latter, but the former shows broader application scenarios. Journal of Chromatography B. https://doi.org/10.1016/j.jchromb.2024.124154
Vancouver
Zhang YY, Guo HL, Wang J, Wang WJ, Li Y, Chu CC, et al. LC-MS/MS and EMIT measure the whole blood concentration of cyclosporine A: The two methods yield concordant results within the dynamic range of the latter, but the former shows broader application scenarios. Journal of Chromatography B. 2024. doi:10.1016/j.jchromb.2024.124154
BibTeX
@article{Zhang2024,
author = {Zhang, Yuan-Yuan and Guo, Hong-Li and Wang, Jie and Wang, Wei-Jun and Li, Yue and Chu, Chen-Chao and Wu, Chun-Ying and Huang, Jian and Hu, Ya-Hui and Chen, Feng},
title = {LC-MS/MS and EMIT measure the whole blood concentration of cyclosporine A: The two methods yield concordant results within the dynamic range of the latter, but the former shows broader application scenarios},
journal = {Journal of Chromatography B},
year = {2024},
doi = {10.1016/j.jchromb.2024.124154}
}
EUVIMED