Energy-integrating detector based ultra-high-resolution CT with deep learning reconstruction for the assessment of calcified lesions in coronary artery disease
Journal of Cardiovascular Computed Tomography · 2024
DOI: 10.1016/j.jcct.2024.09.014 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Journal of Cardiovascular Computed Tomography, 2024. Verfasst von Misato Sone und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Iwate Medical University, Yahaba, Japan. Hauptthemen laut MEDLINE: Coronary Artery Disease, Vascular Calcification, Deep Learning, Phantoms, Imaging. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 30 Literaturangaben. Gefördert von: Japan Society for the Promotion of Science.
Abstract (Originalsprache)
BACKGROUND: The aim of this study to compare of the image quality of calcified lesions in coronary artery disease between deep learning reconstruction (DLR) and model-based iterative reconstruction (MBIR) on energy-integrating detector (EID) based ultra-high-resolution CT (UHRCT). METHODS: We performed a phantom study on EID-based UHRCT using a dedicated insert for calcifications and obtained the derivative values for DLR and MBIR. In the clinical study, the derivative values were compared between DLR and MBIR across 73 calcified lesions in 62 patients. Edge sharpness of calcifications and contrast resolution at the coronary lumen side were quantified by the maximum and minimum derivative values. Two radiologists independently analyzed image quality of the calcified lesions using a 5-point Likert scale. RESULTS: In the phantom study, the edge sharpness of the 3-mm calcifications on DLR (median, 924 HU/mm; IQR, 580-1741 HU/mm) was significantly higher than on MBIR (median, 835 HU/mm; IQR, 484-1552; p < 0.001). In the clinical study, the image quality of the calcified lesions was significantly better on DLR with significantly reduced reconstruction time (p < 0.001). The contrast resolution at the coronary lumen side on DLR (median, -99.1 HU/mm; IQR, -209 to -34.3 HU/mm) was significantly higher than on MBIR (median, -41.8 HU/mm; IQR, -121 to 22.3 HU/mm, p < 0.001) although the edge sharpness of calcifications was similar between DLR and MBIR (p = 0.794) in the clinical setting. CONCLUSION: EID-based UHRCT reconstructed using DLR represents better image quality of calcified lesions in coronary artery disease compared with MBIR, with significantly reduced reconstruction time.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology, Iwate Medical University, Yahaba, Japan.
- Department of Radiology, Iwate Medical University, Yahaba, Japan. Electronic address: kori931@gmail.com.
- Center for Radiological Science, Iwate Medical University, Yahaba, Japan.
- Canon Medical Systems Europe, Amstelveen, Netherlands.
- Canon Medical Systems, Otawara, Japan.
Förderung
- Japan Society for the Promotion of Science
Zitieren
APA
Sone, M., Orii, M., Ota, Y., Chiba, T., Schuijf, J. D., Akino, N., & Yoshioka, K. (2024). Energy-integrating detector based ultra-high-resolution CT with deep learning reconstruction for the assessment of calcified lesions in coronary artery disease. Journal of Cardiovascular Computed Tomography. https://doi.org/10.1016/j.jcct.2024.09.014
Vancouver
Sone M, Orii M, Ota Y, Chiba T, Schuijf JD, Akino N, et al. Energy-integrating detector based ultra-high-resolution CT with deep learning reconstruction for the assessment of calcified lesions in coronary artery disease. Journal of Cardiovascular Computed Tomography. 2024. doi:10.1016/j.jcct.2024.09.014
BibTeX
@article{Sone2024,
author = {Sone, Misato and Orii, Makoto and Ota, Yoshitaka and Chiba, Takuya and Schuijf, Joanne D and Akino, Naruomi and Yoshioka, Kunihiro},
title = {Energy-integrating detector based ultra-high-resolution CT with deep learning reconstruction for the assessment of calcified lesions in coronary artery disease},
journal = {Journal of Cardiovascular Computed Tomography},
year = {2024},
doi = {10.1016/j.jcct.2024.09.014}
}