Test-retest reliability of MyotonPRO parameters in healthy muscles at rest: A profile-based analysis

Flávia Alves de Carvalho, Fernanda Pegorin Diniz, Lucas Antonio Buara Arminio, Isaque Machado da Silva, Carlos Marcelo Pastre

Journal of Bodywork and Movement Therapies · 2026-10 · 48 · 1739-1744

DOI: 10.1016/j.jbmt.2026.09.025 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Bodywork and Movement Therapies, 2026-10. Verfasst von Flávia Alves de Carvalho und 4 weiteren Autor:innen. Die Arbeit nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Conselho Nacional de Desenvolvimento Científico e Tecnológico.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Reliable myotonometric reference values are needed for different muscles and populations to improve the interpretation of these parameters.

Methods

Eighty healthy male and female adults were assessed twice within 48 hours using the MyotonPRO on the dominant middle deltoid (MD), rectus femoris (RF), tibialis anterior (TA), and soleus (SO). Muscle tone, stiffness, elasticity, relaxation time, and creep were analyzed. Reliability was assessed using intraclass correlation coefficients (ICC 3,1) and minimal detectable change (MDC).

Results

Tone, stiffness, and relaxation time showed good reliability in the MD (ICC = 0.84, 0.81, 0.77) and RF (ICC = 0.70, 0.67, 0.78). TA and SO showed poor-to-moderate reliability for tone (ICC = 0.49 and 0.30), stiffness (ICC = 0.43 and 0.13), and relaxation time (ICC = 0.59 and 0.26). Elasticity and creep had limited reliability across all muscles. In women, muscle tone and elasticity were lower, and relaxation time higher, while in men creep was lower.

Conclusions

The MyotonPRO is reliable for assessing tone, stiffness, and relaxation time in the MD and RF. Its use in other muscles or parameters should be approached with caution. Sex differences should be considered in interpretation.

Förderung

  • Conselho Nacional de Desenvolvimento Científico e Tecnológico

Zitieren

APA

Carvalho, F. A. D., Diniz, F. P., Arminio, L. A. B., Silva, I. M. D., & Pastre, C. M. (2026). Test-retest reliability of MyotonPRO parameters in healthy muscles at rest: A profile-based analysis. Journal of Bodywork and Movement Therapies, 48, 1739-1744. https://doi.org/10.1016/j.jbmt.2026.09.025

Vancouver

Carvalho FAD, Diniz FP, Arminio LAB, Silva IMD, Pastre CM. Test-retest reliability of MyotonPRO parameters in healthy muscles at rest: A profile-based analysis. Journal of Bodywork and Movement Therapies. 2026;48:1739-1744. doi:10.1016/j.jbmt.2026.09.025

BibTeX
@article{Carvalho2026,
  author = {Carvalho, Flávia Alves de and Diniz, Fernanda Pegorin and Arminio, Lucas Antonio Buara and Silva, Isaque Machado da and Pastre, Carlos Marcelo},
  title = {Test-retest reliability of MyotonPRO parameters in healthy muscles at rest: A profile-based analysis},
  journal = {Journal of Bodywork and Movement Therapies},
  year = {2026},
  volume = {48},
  pages = {1739--1744},
  doi = {10.1016/j.jbmt.2026.09.025}
}
Kostenlos starten →